background image

Subject to change without notice

10

With a time coefficient of 10 ns/div (X x10 magnification active),
the example shown in the above figure results in a total measured
risetime of

t

tot

 = 1.6 div x 10 ns/div = 16 ns

When very fast risetimes are being measured, the risetimes of
the oscilloscope amplifier and of the attenuator probe have to be
deducted from the measured time value. The risetime of the
signal can be calculated using the following formula.

t

r

 = 

 t

tot2

 – t

osc2

 – t

p2

In this t

tot

 is the total measured risetime, t

osc

 is the risetime of the

oscilloscope amplifier (approx. 7 ns), and t

p

 the risetime of the

probe (e.g. = 2 ns). If t

tot

 is greater than 100 ns, then t

tot

 can be

taken as the risetime of the pulse, and calculation is unnecessary.

Calculation of the example in the figure above results in a signal
risetime:

t

r

 = 

16

2

 – 7

2

 – 2

= 14.25 ns

The measurement of the rise or fall time is not limited to the trace
dimensions shown in the above diagram. It is only particularly
simple in this way. In principle it is possible to measure in any
display position and at any signal amplitude. It is only important
that the full height of the signal edge of interest is visible in its full
length at not too great steepness and that the horizontal distance
at 10% and 90% of the amplitude is measured. If the edge shows
rounding or overshooting, the 100% should not be related to the
peak values but to the mean pulse heights. Breaks or peaks
(glitches) next to the edge are also not taken into account. With
very severe transient distortions, the rise and fall time
measurement has little meaning. For amplifiers with approximately
constant group delay (therefore good pulse transmission
performance) the following numerical relationship between rise
time tr (in ns) and bandwidth B (in MHz) applies:

Connection of Test Signal

In most cases, briefly depressing the

 AUTOSET 

causes a useful

signal related instrument setting. The following explanations
refer to special applications and/or signals, demanding a manual
instrument setting.

 The description of the controls is explained

in the section ”controls and readout”

.

Caution:
When connecting unknown signals to the oscilloscope input,
always use automatic triggering and set the input coup-
ling switch to AC. The attenuator should initially be set to
20 V/div.

Sometimes the trace will disappear after an input signal has been
applied. Then a higher deflection coefficient (lower input sensitivity)
must be chosen until the vertical signal height is only 3 – 8 div.
With a signal amplitude greater than 160 V

pp

 and the deflection

coefficient (

VOLTS/DIV.

) in calibrated condition, an attenuator

probe must be inserted before the Y input. If, after applying the
signal, the trace is nearly blanked, the period of the signal is
probably substantially longer than the set time deflection
coefficient (

TIME/DIV.

). It should be switched to an adequately

larger time coefficient.

Examples:

Displayed wavelength L = 7 div,
set time coefficient Tc = 100 ns/div,
thus period T = 7 x 100 x 10

-9

 = 0.7 µs

thus freq. F = 1/(0.7 x 10

-6

) = 1.428 MHz.

Signal period T = 1s,
set time coefficient Tc = 0.2 s/div,
thus wavelength L = 1/0.2 = 5 div.

Displayed ripple wavelength L = 1 div,
set time coefficient Tc = 10 ms/div,
thus ripple freq. F = 1/(1 x 10 x 10

-3

) = 100 Hz.

TV Line frequency F = 15625 Hz,
set time coefficient Tc = 10 µs/div,
required wavelength L = 1/(15,625 x 10

-5

) = 6.4 div.

Sine wavelength L = min. 4 div, max. 10 div,
Frequency F = 1 kHz,
max. time coefficient Tc = 1/(4 x 10

3

) = 0.25 ms/div,

min. time coefficient Tc = 1/(10 x 10

3

) = 0.1 ms/div,

set time coefficient Tc = 0.2 ms/div,
required wavelength L = 1/(10

x 0.2 x 10

-3

) = 5 div.

Displayed wavelength L = 0.8 div,
set time coefficient Tc = 0.5 µs/div,
pressed X-MAG. (x10) button: Tc = 0.05 µs/div,
thus  freq. F = 1/(0.8 x 0.05 x 10

-6

) = 25 MHz,

thus period T = 1/(25 x 10

6

) = 40 ns.

If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied

(X-MAG. (x10)

 active). In this case, the time interval of interest

can be shifted to the screen center using the

 X-POS. 

control.

Rise Time  Measurement

When investigating pulse or square waveforms, the critical
feature is the rise time of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly influence
the measuring accuracy, the rise time is generally measured
between 10% and 90% of the vertical pulse height. For
measurement, adjust the Y deflection coefficient using its variab-
le function (uncalibrated) together with the Y-POS. control so that
the pulse height is precisely aligned with the 0% and 100% lines
of the internal graticule. The 10% and 90% points of the signal
will now coincide with the 10% and 90% graticule lines. The
risetime is given by the product of the horizontal distance in div
between these two coincident points and the calibrated time
coefficient setting. The fall time of a pulse can also be measured
by using this method.
The following figure shows correct positioning of the oscilloscope
trace for accurate rise time measurement.

Type of signal voltage

Summary of Contents for HM504-2

Page 1: ...furbished and purchasing new Financing options such as Financial Rental and Leasing are also available on application TMG will assist if you are unsure whether this model will suit your requirements C...

Page 2: ......

Page 3: ...O s c i l l o s c o p e H M 5 0 4 2 Manual English...

Page 4: ...25 Phase comparison with Lissajous figures 26 Phase difference measurement in DUAL mode Yt 26 Phase difference measurement in DUAL mode 26 Measurement of an amplitude modulation 26 Triggering and tim...

Page 5: ...trol cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the in terior of the oscilloscope is screened by the cabinet direct radiation can...

Page 6: ...signal level Time Base delay provide high X Magnification of any portion of the signal 100 MHz 4 Digit Frequency Counter Cursor and Automatic Measurement Save Recall Memories for Instrument Settings R...

Page 7: ...ontrols Autoset automatic signal related parameter settings Save and Recall 9 instrument parameter settings Readout display of menu parameters cursors and results Autom Measurement Freq Period Vdc Vpp...

Page 8: ...the handle can be set into diffe rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked STOP Attention When changing t...

Page 9: ...nrequires the horizontal or inclined position STOP Do not obstruct the ventilation holes Specifications are valid after a 20 minute warm up period between 15 and 30 degr C Specifications without toler...

Page 10: ...pprox 10 and the real voltage value is 11 higher The gain reduction error can not be defined exactly as the 3 dB bandwidth of the Y amplifiers differs between 50 MHz and 55 MHz When examining square o...

Page 11: ...pe However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 2...

Page 12: ...nt lowerinputsensitivity must be chosen until the vertical signal height is only 3 8 div With a signal amplitude greater than 160 Vpp and the deflection coefficient VOLTS DIV in calibrated condition a...

Page 13: ...eristic resistance of the cable as a rule 50 Ohm Especially when transmitting square and pulse signals a resistor equal to the characteristic impedance of the cable must also be connected across the c...

Page 14: ...d by reducing the READOUT intensity RO The INTENS FOCUS control knob adjusts the READOUT intensity Turning this knob clockwise increases and counter clockwisedecreasestheintensity Onlytheminimumrequir...

Page 15: ...Rot with Int trace rotation by use of INTENS FOCUS knob and allows you to compensate the influence of the Earth s magneticfieldonthetracedeflection Toavoidmisadjustment due to unavoidable deflection...

Page 16: ...selected This allows for a different slope setting for the triggered DELAY DTR time base mode 10 TR Trigger indicator LED The TR LED is lit in Yt mode if the triggering conditions are met for the firs...

Page 17: ...the readout e g Y1 deflection coefficient input coupling The symbolizes calibrated measuring conditions and is replaced by the symbol in uncalibrated conditions VAR The vernier variable function is de...

Page 18: ...e internal trigger signal originating from channel I or channel II or the external trigger signal can be chosen DUAL and Addition mode Briefly pressing opens the trigger source Pulldown menu with the...

Page 19: ...s the way the trigger signal is connected to the trigger unit AC DC content suppressed DC peak value detection inactive HF high pass filter cuts off frequencies below approx 50 kHz trigger point symbo...

Page 20: ...e trace start position which is present after switching over from sea to del This enables lower time deflection coefficient settings for signal expansion del In DELAY mode a trigger event does not sta...

Page 21: ...PUT CH II BNC socket This BNC socket is the signal input for channel II The outer ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of t...

Page 22: ...ifferent frequency responses of the trigger amplifier and Y amplifier cause a reduction of the measurement accuracy If relatively low frequency signals 20 Hz are present the measurement value continuo...

Page 23: ...io X measurement causes the display of two long and one short CURSOR lines and is enabled inYt time base mode only The unit to be displayed must be selected by briefly pressing the UNIT 35 pushbutton...

Page 24: ...es are displayed Briefly pressing SOURCE selects the channel and it s deflection coefficient for the measurement The CURSOR lines must be set to the signal according to the selected channel b DUAL mod...

Page 25: ...ubmenus Main Menue 50 Ohm For high impedance loads Oscilloscope approx 1 M Ohm Digital Voltmeter approx 10 MOhm the output voltage is either 0 2 Volt DC or 0 2 Vpp AC square wave Under First Time Oper...

Page 26: ...n generator provides a square wave signal with selectable frequencies and a very fast rise time 4 ns from the output socket below the CRT screen As the square wave signals are used for probe compensat...

Page 27: ...the calibrator output via a suitable probe The voltage provided by the probe to a high impedance input 1 MOhm II 15 30 pF will correspond to the division ratio of the probe used 10 1 20 mVpp output S...

Page 28: ...the phosphor at too high a brightness setting INTENS setting which causes either a lasting loss of brightness or in the extreme case complete destruction of the phosphor at this point Phase difference...

Page 29: ...a sine signal Except when external trigger is used the trigger threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger...

Page 30: ...formation regarding this item can be noted in the Data Sheet The coupling setting TRIG MODE 20 and indication are described under Controls and Readout As the automatic triggering does not work below 2...

Page 31: ...er threshold less than 0 5 div It is therefore particularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in a circuit In this trigger mode the slope di...

Page 32: ...gnals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use of the holdoff control simplifies t...

Page 33: ...osen in del and dTr mode is stored and automatically set after activating one of those modes If the stored time coefficient in del dTr mode was higher than the actual value in sea search mode the time...

Page 34: ...me base mode I X x10 magnifier switched off I Optimum X and Y position settings I Trace and readout visible If DC trigger coupling had been selected AC trigger coupling will not be chosen and the auto...

Page 35: ...w typical patterns displayed by the various components under test I Open circuit is indicated by a straight horizontal line I Short circuit is shown by a straight vertical line Testing Resistors If th...

Page 36: ...or ground for the same terminal is then absolutely necessary A connection inversion effects a rotation of the test pattern by 180 degrees about the centre point of the scope graticule Pay attention t...

Page 37: ...ation The oscilloscope is supplied with a serial interface for control purposes The interface connector 9 pole D SUB female is located on the rear of the instrument Via this bidirectional port the ins...

Page 38: ...RECALL INPUT CHI X AC DC GND max 400Vp 1M II 18pF INPUT CHI X max 400Vp 1M II 18pF TRIG EXT max 100Vp INP Z x1 x10 AC DC GND x1 x10 50 MHz ANALOG OSCILLOSCOPE HM504 2 Instruments CAT I TRACE ROT CT GL...

Page 39: ...37 Subject to change without notice Front Panel HM504 2...

Page 40: ......

Page 41: ...LEER...

Page 42: ...chwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes Spectrum Analyzer P...

Reviews: