background image

Subject to change without notice

28

The automatic peak (value) triggering operates over all variations
or fluctuations of the test signal above 20 Hz. However, if the
pulse duty factor of a square wave signal exceeds a ratio of
100 : 1, switching over to normal triggering will be necessary.
Automatic triggering is practicable with internal and external
trigger voltage.

Normal Triggering

Information specific to the instrument is given in the sections 

NM

- AT -  [9], LEVEL [11] 

and TRIG. MODE 

[20] 

in the paragraphs

”Controls and Readout”. The time fine adjustment  (VAR.), and
the holdoff time setting assist in triggering under specially difficult
signal conditions.

With normal triggering, the sweep can be started by

 AC

 signals

within the frequency range defined by the trigger coupling setting.
In the absence of an adequate trigger signal or when the trigger
controls (particularly the trigger

 LEVEL 

control) are misadjusted,

no trace is visible.

When using the internal normal triggering mode, it is possible to
trigger at any amplitude point of a signal edge, even with very
complex signal shapes, by adjusting the trigger

 LEVEL

 control. If

the signal applied at the Y input is used for triggering (internal
trigger source), its adjusting range is directly dependent on the
display height, which should be at least 0.5div. If it is smaller than
1div, the trigger

 LEVEL 

adjustment needs to be operated with a

sensitive touch. In the external normal triggering mode, the same
applies to approx. 0.3 V

pp

 external trigger voltage amplitude.

Other measures for triggering of very complex signals are the use
of the time base variable control and

 HOLDOFF 

time control,

mentioned below.

SLOPE 

Please note item [9] in section ”Controls and Readout” for
instrument specific information.

The actual slope setting is displayed in the readout. The setting
is not changed by the

 AUTO SET 

function. The slope setting can

be changed for the delay time base trigger unit in delay mode if
the delay trigger function is active. The previous slope setting for
the undelayed time base trigger is stored and still active. For
further information please note ”Controls and Readout”.

The time base generator can be triggered by a rising or falling
edge of the test signal. Whether the rising or the falling edge is
used for triggering, depends on the slope direction setting. This
is valid with automatic and normal triggering. The positive slope
direction means an edge going from a negative potential and
rising to a positive potential. This has nothing to do with zero or
ground potential or absolute voltage values. The positive slope
may also lie in a negative part of a signal.

However the trigger point may be varied within certain limits on
the chosen edge using the

 LEVEL 

control. The slope direction is

always related to the input signal and the non inverted display.

Trigger coupling

Instrument specific information regarding this item can be noted
in the ”Data Sheet”. The coupling setting 

TRIG. MODE [20]

 and

indication are described under ”Controls and Readout”.

As the automatic triggering does not work below 20 Hz, normal
triggering should be used in DC and LF trigger coupling mode. The
coupling mode and accordingly the frequency range of the trigger
signal should meet the signal requirements.

AC:

This is the most frequently used trigger mode. The trigger
threshold increases below and above the frequency limits
mentioned in the data sheet. This filter cuts off both the DC
content of the trigger signal and the lowest frequency range.

DC:

In this coupling mode the trigger signal is coupled galvanically
to the trigger unit if normal triggering (NM) is present.
Therefore there is no low frequency limit.

DC triggering is recommended if the signal is to be triggered
with quite slow processes or if pulse signals with constantly
changing pulse duty factors have to be displayed.

HF:

In this coupling mode the transmission range equals a high
pass filter. It cuts off the DC content of the trigger signal and
the lower frequency range.

LF:

LF trigger coupling has a low pass filter function characteristic.
As in DC trigger coupling, there is no limit for the pass
frequency range in connection with normal triggering.

The LF trigger coupling is often more suitable for low frequency
signals than DC trigger coupling because the noise
components of the trigger signals are strongly suppressed.
This avoids or reduces, under borderline conditions, jitter or
double traces especially with very low signal voltages. The
trigger threshold rises continuously above the pass band.

TVL:

 The built in active TV Sync Separator provides the separation

of line sync pulses from the video signal.

Even distorted video signals are triggered and displayed in a
stable manner. This mode is described under paragraph
”Triggering of video signals”.

TVF:

 The built in active TV Sync Separator also provides the

separation of frame sync pulses from the video signal. Even
distorted video signals are triggered and displayed in a stable
manner.

 This mode is described under paragraph ”Triggering

of video signals”.

~:

Indicates “line/mains triggering” and is described under the
paragraph of the same name.

Triggering of video signals

In

 TVL 

and

 TVF 

trigger coupling mode the instrument is

automatically set to automatic triggering and the trigger point
indicator is switched off. As only the separated synchronization
pulses are used for triggering the relationship between the
displayed signal and the trigger signal is lost. In

 TVF 

mode

interference may occur if chopped

 DUAL 

mode is chosen or the

readout is active.

Video signals are triggered in the automatic mode. The internal
triggering is virtually independent of the display height, but the
sync pulse must exceed 0.5div height.

The polarity of the synchronization pulse is critical for the slope
selection. If the displayed sync pulses are above the picture (field)
contents (leading edge positive going), then the slope setting for

Triggering and time base

Summary of Contents for HM504-2

Page 1: ...furbished and purchasing new Financing options such as Financial Rental and Leasing are also available on application TMG will assist if you are unsure whether this model will suit your requirements C...

Page 2: ......

Page 3: ...O s c i l l o s c o p e H M 5 0 4 2 Manual English...

Page 4: ...25 Phase comparison with Lissajous figures 26 Phase difference measurement in DUAL mode Yt 26 Phase difference measurement in DUAL mode 26 Measurement of an amplitude modulation 26 Triggering and tim...

Page 5: ...trol cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the in terior of the oscilloscope is screened by the cabinet direct radiation can...

Page 6: ...signal level Time Base delay provide high X Magnification of any portion of the signal 100 MHz 4 Digit Frequency Counter Cursor and Automatic Measurement Save Recall Memories for Instrument Settings R...

Page 7: ...ontrols Autoset automatic signal related parameter settings Save and Recall 9 instrument parameter settings Readout display of menu parameters cursors and results Autom Measurement Freq Period Vdc Vpp...

Page 8: ...the handle can be set into diffe rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked STOP Attention When changing t...

Page 9: ...nrequires the horizontal or inclined position STOP Do not obstruct the ventilation holes Specifications are valid after a 20 minute warm up period between 15 and 30 degr C Specifications without toler...

Page 10: ...pprox 10 and the real voltage value is 11 higher The gain reduction error can not be defined exactly as the 3 dB bandwidth of the Y amplifiers differs between 50 MHz and 55 MHz When examining square o...

Page 11: ...pe However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 2...

Page 12: ...nt lowerinputsensitivity must be chosen until the vertical signal height is only 3 8 div With a signal amplitude greater than 160 Vpp and the deflection coefficient VOLTS DIV in calibrated condition a...

Page 13: ...eristic resistance of the cable as a rule 50 Ohm Especially when transmitting square and pulse signals a resistor equal to the characteristic impedance of the cable must also be connected across the c...

Page 14: ...d by reducing the READOUT intensity RO The INTENS FOCUS control knob adjusts the READOUT intensity Turning this knob clockwise increases and counter clockwisedecreasestheintensity Onlytheminimumrequir...

Page 15: ...Rot with Int trace rotation by use of INTENS FOCUS knob and allows you to compensate the influence of the Earth s magneticfieldonthetracedeflection Toavoidmisadjustment due to unavoidable deflection...

Page 16: ...selected This allows for a different slope setting for the triggered DELAY DTR time base mode 10 TR Trigger indicator LED The TR LED is lit in Yt mode if the triggering conditions are met for the firs...

Page 17: ...the readout e g Y1 deflection coefficient input coupling The symbolizes calibrated measuring conditions and is replaced by the symbol in uncalibrated conditions VAR The vernier variable function is de...

Page 18: ...e internal trigger signal originating from channel I or channel II or the external trigger signal can be chosen DUAL and Addition mode Briefly pressing opens the trigger source Pulldown menu with the...

Page 19: ...s the way the trigger signal is connected to the trigger unit AC DC content suppressed DC peak value detection inactive HF high pass filter cuts off frequencies below approx 50 kHz trigger point symbo...

Page 20: ...e trace start position which is present after switching over from sea to del This enables lower time deflection coefficient settings for signal expansion del In DELAY mode a trigger event does not sta...

Page 21: ...PUT CH II BNC socket This BNC socket is the signal input for channel II The outer ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of t...

Page 22: ...ifferent frequency responses of the trigger amplifier and Y amplifier cause a reduction of the measurement accuracy If relatively low frequency signals 20 Hz are present the measurement value continuo...

Page 23: ...io X measurement causes the display of two long and one short CURSOR lines and is enabled inYt time base mode only The unit to be displayed must be selected by briefly pressing the UNIT 35 pushbutton...

Page 24: ...es are displayed Briefly pressing SOURCE selects the channel and it s deflection coefficient for the measurement The CURSOR lines must be set to the signal according to the selected channel b DUAL mod...

Page 25: ...ubmenus Main Menue 50 Ohm For high impedance loads Oscilloscope approx 1 M Ohm Digital Voltmeter approx 10 MOhm the output voltage is either 0 2 Volt DC or 0 2 Vpp AC square wave Under First Time Oper...

Page 26: ...n generator provides a square wave signal with selectable frequencies and a very fast rise time 4 ns from the output socket below the CRT screen As the square wave signals are used for probe compensat...

Page 27: ...the calibrator output via a suitable probe The voltage provided by the probe to a high impedance input 1 MOhm II 15 30 pF will correspond to the division ratio of the probe used 10 1 20 mVpp output S...

Page 28: ...the phosphor at too high a brightness setting INTENS setting which causes either a lasting loss of brightness or in the extreme case complete destruction of the phosphor at this point Phase difference...

Page 29: ...a sine signal Except when external trigger is used the trigger threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger...

Page 30: ...formation regarding this item can be noted in the Data Sheet The coupling setting TRIG MODE 20 and indication are described under Controls and Readout As the automatic triggering does not work below 2...

Page 31: ...er threshold less than 0 5 div It is therefore particularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in a circuit In this trigger mode the slope di...

Page 32: ...gnals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use of the holdoff control simplifies t...

Page 33: ...osen in del and dTr mode is stored and automatically set after activating one of those modes If the stored time coefficient in del dTr mode was higher than the actual value in sea search mode the time...

Page 34: ...me base mode I X x10 magnifier switched off I Optimum X and Y position settings I Trace and readout visible If DC trigger coupling had been selected AC trigger coupling will not be chosen and the auto...

Page 35: ...w typical patterns displayed by the various components under test I Open circuit is indicated by a straight horizontal line I Short circuit is shown by a straight vertical line Testing Resistors If th...

Page 36: ...or ground for the same terminal is then absolutely necessary A connection inversion effects a rotation of the test pattern by 180 degrees about the centre point of the scope graticule Pay attention t...

Page 37: ...ation The oscilloscope is supplied with a serial interface for control purposes The interface connector 9 pole D SUB female is located on the rear of the instrument Via this bidirectional port the ins...

Page 38: ...RECALL INPUT CHI X AC DC GND max 400Vp 1M II 18pF INPUT CHI X max 400Vp 1M II 18pF TRIG EXT max 100Vp INP Z x1 x10 AC DC GND x1 x10 50 MHz ANALOG OSCILLOSCOPE HM504 2 Instruments CAT I TRACE ROT CT GL...

Page 39: ...37 Subject to change without notice Front Panel HM504 2...

Page 40: ......

Page 41: ...LEER...

Page 42: ...chwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes Spectrum Analyzer P...

Reviews: