background image

21

Subject to change without notice

1/

∆∆∆∆∆

(display “1/

t: measured value“)

Two vertical CURSOR lines enable frequency measurement
in Yt mode (not in XY mode). Briefly pressing the UNIT 

[35]

pushbutton directly switches over from 

1/

∆∆∆∆∆

to

 

∆∆∆∆∆

(time)

measurement and vice versa.

Rise Time

 (display “tr 10: measured value”)

Rise time measurement by aid of two horizontal CURSOR
lines and two “+” symbols which have the following meaning.
1

st

Lower CURSOR line = 0%.

2

nd

Lower “+” symbol = 10% of the CURSOR lines
distance.

3

rd

Upper “+” symbol = 90% of the CURSOR lines
distance.

4

th

Upper CURSOR line = 100%.

SET

 

[32]

 enables an automatic signal related CURSOR line

setting (in DUAL mode related to the signal used for triggering),
which can later be changed manually.
The distance between the “+” symbols and the CURSOR
lines are set automatically. For rise time measurement the
horizontal position of the “+” symbols must be set manually
to the signal slope. This requires that the CURSOR POS is
active and each “+” symbol is marked (“I”, “II”) by the aid
of the SELECT 

[34]

 function.

Note:

 For maximum “+” symbol positioning and measuring

accuracy first set the signal slope to the screen centre X-
POS. 

[12]

  and then activate X-magnifier X-MAG. x10 

[13].

To avoid CURSOR line and “+” symbol changes after each
change of a signal position  in X and/or Y direction, a fixed
relation between signal and CURSOR display can be made by
activating the GLUE (33) function. GLUE is indicated by a
reduced number of dots in the CURSOR lines and the “+”
symbols.
Further information about this item can be found in this
manual under “Type of signal voltage” in section “Rise Time
Measurement”.

∆∆∆∆∆

V

 (display 

V: channel, measured value)

CURSOR supported voltage measurement.
In Yt (time base) mode two horizontal CURSOR lines are
displayed: Single channel (CH I or CH II) mode automatically
relates one signal and the CURSOR lines. The measurement
value is connected with the Y deflection coefficient.
DUAL mode requires selection between channel I and II with
the SOURCE -pushbutton 

[33]

. The CURSOR line must be

placed on the signal (channel) chosen by the SOURCE
function.
Addition (“add”) mode requires equal Y deflection coefficients
for both channels.
XY mode causes the display of two vertical or horizontal
CURSOR lines: The SOURCE-pushbutton 

[33]

 allows selection

between X (CH I) and Y (CH II) voltage measurement. In the
case of X voltage measurement, vertical CURSOR lines are
displayed.

V to GND

 (display V: channel, measured value)

One CURSOR line is displayed for voltage measurement
related to the trace 0 Volt position. This is the only exception
to the description of item 

V.

Ratio X

 (display “ratio:X, measured value, unit”)

Ratio X measurement causes the display of two long and one
short CURSOR lines and is enabled in Yt (time base) mode only.

The unit to be displayed must be selected by briefly pressing
the UNIT 

[35]

 pushbutton to call the UNIT menu. Then the

following units are being offered: ratio, %, ° (angle unit:
degree of angle and pi.
The long CURSOR line in the left position always serves as
reference line. A “–“ (minus) sign indicates measurement
results if the short CURSOR line is placed left of the reference
line.

Ratio:

Enables the measurement of pulse duty ratio. The distance
between the long CURSOR lines is equivalent to1 (whole
cycle).
Example for a pulse signal with 4 div. pulse and 1 div. space:
The long CURSOR lines must coincide with the start position
of first and the second pulse (distance = 5 div.) as the
reference distance (1). Then the “I” symbol must be assigned
to the short CURSOR line SELECT 

[34]

 which must then be

set to the pulse end position (4 div. after the pulse start).
Corresponding to the ratio of pulse duration to period length
(4:5 = 0.8) “0.8” will be displayed.

%:

Same function as described before under “Ratio”. The
measurement result is displayed in % (unit).

°

:

Angle measurement referring to the CURSOR line distances.
The distance between the long CURSOR lines should cover
one signal period, equivalent to 360°. Angle measurement
then can be performed by shifting the short CURSOR line.
Additional information can be found in section “Operating
modes of the vertical amplifiers in Yt mode” under “Phase
difference measurement in DUAL mode”.

pi:

Determination of the value for “pi” referring to the CURSOR
line distances. The equivalent for “2 pi” is one sine wave
period; thus the distance between the long CURSOR lines
must be 1 period. If the distance between the long CURSOR
in left hand position and the short CURSOR line referring to
it is 1.5 periods, “3 pi” will be indicated.

Ratio Y

 (display “ratio:Y, measured value, unit”)

Ratio Y measurement causes the display of two long and one
short CURSOR lines and is enabled only in Yt (time base)
mode.
Briefly pressing the UNIT-pushbutton 

[35]

 switches over

between the ratio (unnamed) and ratio in %.
The long CURSOR line in the lower position always serves as
the reference line. A “–“ (minus) sign indicates measurement
results if the short CURSOR line is placed below  the
reference line.

Ratio:

 The distance between both long CURSOR lines is

equivalent to1.
Example: If the distance between the long CURSOR lines is
6 div. and the short CURSOR line is activated SELECT 

[34]

and set 4 div. above the reference CURSOR line, the ratio is
4:6, causing “0.667” (without unit) to be displayed.

%:

The only difference between previous item “Ratio” and “%”
is that the distance between the long CURSOR lines is
equivalent to100% and the measuring result is displayed as
a % value.

Controls and Readout

Summary of Contents for HM504-2

Page 1: ...furbished and purchasing new Financing options such as Financial Rental and Leasing are also available on application TMG will assist if you are unsure whether this model will suit your requirements C...

Page 2: ......

Page 3: ...O s c i l l o s c o p e H M 5 0 4 2 Manual English...

Page 4: ...25 Phase comparison with Lissajous figures 26 Phase difference measurement in DUAL mode Yt 26 Phase difference measurement in DUAL mode 26 Measurement of an amplitude modulation 26 Triggering and tim...

Page 5: ...trol cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the in terior of the oscilloscope is screened by the cabinet direct radiation can...

Page 6: ...signal level Time Base delay provide high X Magnification of any portion of the signal 100 MHz 4 Digit Frequency Counter Cursor and Automatic Measurement Save Recall Memories for Instrument Settings R...

Page 7: ...ontrols Autoset automatic signal related parameter settings Save and Recall 9 instrument parameter settings Readout display of menu parameters cursors and results Autom Measurement Freq Period Vdc Vpp...

Page 8: ...the handle can be set into diffe rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked STOP Attention When changing t...

Page 9: ...nrequires the horizontal or inclined position STOP Do not obstruct the ventilation holes Specifications are valid after a 20 minute warm up period between 15 and 30 degr C Specifications without toler...

Page 10: ...pprox 10 and the real voltage value is 11 higher The gain reduction error can not be defined exactly as the 3 dB bandwidth of the Y amplifiers differs between 50 MHz and 55 MHz When examining square o...

Page 11: ...pe However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 2...

Page 12: ...nt lowerinputsensitivity must be chosen until the vertical signal height is only 3 8 div With a signal amplitude greater than 160 Vpp and the deflection coefficient VOLTS DIV in calibrated condition a...

Page 13: ...eristic resistance of the cable as a rule 50 Ohm Especially when transmitting square and pulse signals a resistor equal to the characteristic impedance of the cable must also be connected across the c...

Page 14: ...d by reducing the READOUT intensity RO The INTENS FOCUS control knob adjusts the READOUT intensity Turning this knob clockwise increases and counter clockwisedecreasestheintensity Onlytheminimumrequir...

Page 15: ...Rot with Int trace rotation by use of INTENS FOCUS knob and allows you to compensate the influence of the Earth s magneticfieldonthetracedeflection Toavoidmisadjustment due to unavoidable deflection...

Page 16: ...selected This allows for a different slope setting for the triggered DELAY DTR time base mode 10 TR Trigger indicator LED The TR LED is lit in Yt mode if the triggering conditions are met for the firs...

Page 17: ...the readout e g Y1 deflection coefficient input coupling The symbolizes calibrated measuring conditions and is replaced by the symbol in uncalibrated conditions VAR The vernier variable function is de...

Page 18: ...e internal trigger signal originating from channel I or channel II or the external trigger signal can be chosen DUAL and Addition mode Briefly pressing opens the trigger source Pulldown menu with the...

Page 19: ...s the way the trigger signal is connected to the trigger unit AC DC content suppressed DC peak value detection inactive HF high pass filter cuts off frequencies below approx 50 kHz trigger point symbo...

Page 20: ...e trace start position which is present after switching over from sea to del This enables lower time deflection coefficient settings for signal expansion del In DELAY mode a trigger event does not sta...

Page 21: ...PUT CH II BNC socket This BNC socket is the signal input for channel II The outer ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of t...

Page 22: ...ifferent frequency responses of the trigger amplifier and Y amplifier cause a reduction of the measurement accuracy If relatively low frequency signals 20 Hz are present the measurement value continuo...

Page 23: ...io X measurement causes the display of two long and one short CURSOR lines and is enabled inYt time base mode only The unit to be displayed must be selected by briefly pressing the UNIT 35 pushbutton...

Page 24: ...es are displayed Briefly pressing SOURCE selects the channel and it s deflection coefficient for the measurement The CURSOR lines must be set to the signal according to the selected channel b DUAL mod...

Page 25: ...ubmenus Main Menue 50 Ohm For high impedance loads Oscilloscope approx 1 M Ohm Digital Voltmeter approx 10 MOhm the output voltage is either 0 2 Volt DC or 0 2 Vpp AC square wave Under First Time Oper...

Page 26: ...n generator provides a square wave signal with selectable frequencies and a very fast rise time 4 ns from the output socket below the CRT screen As the square wave signals are used for probe compensat...

Page 27: ...the calibrator output via a suitable probe The voltage provided by the probe to a high impedance input 1 MOhm II 15 30 pF will correspond to the division ratio of the probe used 10 1 20 mVpp output S...

Page 28: ...the phosphor at too high a brightness setting INTENS setting which causes either a lasting loss of brightness or in the extreme case complete destruction of the phosphor at this point Phase difference...

Page 29: ...a sine signal Except when external trigger is used the trigger threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger...

Page 30: ...formation regarding this item can be noted in the Data Sheet The coupling setting TRIG MODE 20 and indication are described under Controls and Readout As the automatic triggering does not work below 2...

Page 31: ...er threshold less than 0 5 div It is therefore particularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in a circuit In this trigger mode the slope di...

Page 32: ...gnals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use of the holdoff control simplifies t...

Page 33: ...osen in del and dTr mode is stored and automatically set after activating one of those modes If the stored time coefficient in del dTr mode was higher than the actual value in sea search mode the time...

Page 34: ...me base mode I X x10 magnifier switched off I Optimum X and Y position settings I Trace and readout visible If DC trigger coupling had been selected AC trigger coupling will not be chosen and the auto...

Page 35: ...w typical patterns displayed by the various components under test I Open circuit is indicated by a straight horizontal line I Short circuit is shown by a straight vertical line Testing Resistors If th...

Page 36: ...or ground for the same terminal is then absolutely necessary A connection inversion effects a rotation of the test pattern by 180 degrees about the centre point of the scope graticule Pay attention t...

Page 37: ...ation The oscilloscope is supplied with a serial interface for control purposes The interface connector 9 pole D SUB female is located on the rear of the instrument Via this bidirectional port the ins...

Page 38: ...RECALL INPUT CHI X AC DC GND max 400Vp 1M II 18pF INPUT CHI X max 400Vp 1M II 18pF TRIG EXT max 100Vp INP Z x1 x10 AC DC GND x1 x10 50 MHz ANALOG OSCILLOSCOPE HM504 2 Instruments CAT I TRACE ROT CT GL...

Page 39: ...37 Subject to change without notice Front Panel HM504 2...

Page 40: ......

Page 41: ...LEER...

Page 42: ...chwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes Spectrum Analyzer P...

Reviews: