background image

Subject to change without notice

32

The manipulation of time delay requires a certain experience,
especially with  complicated signal combinations which are
difficult to display. The display of sections of simple signals  is, in
contrast, fairly easy. The time delayed display is also possible in
the dual channel, addition and difference modes.

In chopped DUAL mode, if after switching over to “del” or “dTr”,
the time deflection coefficient is reduced (TIME/DIV.), the channel
switching mode doesn`t change automatically to alternate.

Attention:
In chopped DUAL mode, using high expansion ratios in “del”
mode, chop interference may be visible. This can be overcome
by selecting alternate DUAL mode. A similar effect can be
caused by the READOUT with the result that parts of a signal
displayed in CH I, CH II or DUAL mode are blanked
(unsynchronised). In such a case the READOUT can be
switched off.

AUTOSET

The instrument specific information regarding this function is part
of the section ”Controls and Readout” paragraph

 AUTOSET [2]

.

As also mentioned in that section, all controls are electronically
selected with the exception of the

 POWER 

pushbutton.

Thus automatic, signal related instrument set up in Yt (time base)
mode is possible. In most cases no additional manual instrument
setting is required.

Briefly pressing the

 AUTOSET 

pushbutton causes the instrument

to switch over to the last Yt mode settings regarding

 CH I

,

 CH II

and

 DUAL

. If the instrument was operated in Yt mode, the actual

setting will not be affected with the exception of 

ADD 

mode

which will be switched off. At the same time the attenuator(s)
(

VOLTS/DIV

) are automatically set for a signal display height of

approx. 6 div in mono channel mode or if in 

DUAL 

mode for

approx. 4 div height for each channel. In the determination of the
time deflection coefficient, it is assumed that the pulse duty
factor of the input signal is approx. 1:1.

The time deflection coefficient is also set automatically for a
display of approx. 2 signal periods. The time base setting occurs
randomly if complex signals consisting several frequencies e.g.
video signals are present. If cursor voltage measurement is
selected, AUTOSET also affects the position of the CURSOR
lines. 

Please note

 AUTOSET [2] 

in section ”Controls and

Readout”.

AUTOSET 

sets the instrument automatically to the following

operating conditions:

I

AC or DC input coupling unaltered
or in GND condition the last used setting

I

Internal triggering (channel I or channel II)

I

Automatic triggering

I

Trigger level in electrical midrange position

I

Optimum calibrated Y deflection coefficient(s) 5 mV - 20 mv/div

I

Optimum calibrated Time base deflection coefficient

I

AC trigger coupling (except if DC trigger coupling last present

I

Undelayed time base mode

I

X x10 magnifier switched off

I

Optimum X and Y position settings

I

Trace and readout visible.

If

 DC 

trigger coupling had been selected,

 AC 

trigger coupling will

not be chosen and the automatic trigger is operative without the
peak value detection.

The

 

X position is set to the CRT centre as well as the Y position

in

 CH I 

or

 CH II 

mode. In

 DUAL 

mode the channel I trace is set to

the upper half and the channel II trace to the lower half of the CRT.

The 1 mV/div and 2 mV/div deflection coefficient will not be
selected by

 AUTOSET 

as the bandwidth is reduced on these

settings.

Attention!
If a signal is applied with a pulse duty factor of approx. 400 :1
or larger, an automatic signal display can not be performed.
The pulse duty factor causes too low a Y deflection coefficient
(sensitivity too high) and too high a time deflection coefficient
(time base speed to slow) and results in a display in which
only the baseline is visible.

In such cases it is recommended to select normal triggering and
to set the trigger point approx. 0.5div above or below the trace.
If under one of these conditions the trigger indicator LED is lit, this
indicates the presence of a signal. Then both the time coefficient
and Y deflection coefficient should be reduced. Please note that
a reduction in intensity may occur, which could result in a blank
screen when the physical limits are reached.

Mean Value Display

The DC Mean Value is displayed in place of the cursor line
measurement, if the cursor lines are switched off, the AUTO
MEASURE menu function “DC” is activated and further condition
are met:
The signal to be measured (AC > 20 Hz) must be applied at input
CH I 

[25]

 or CH II 

[28]

 with its DC content at the measuring

amplifier; DC input coupling 

[26; 29]

 required. Yt (time base) mode

in combination with internal triggering (trigger source CH I or
CH II; not alternated triggering) must be present. AC or DC trigger
coupling must be selected.
If the above conditions are not met, ”n/a” will be displayed.

The mean value is acquired using the trigger signal amplifiers
used for internal triggering. With the exception of DUAL mode,
the indicated mean value is automatically related to the active
channel (CH I or CH II), as the channel selection also assigns the
trigger amplifier. In DUAL mode one can select between trigger
amplifier CH I or CH II for triggering. The indicated mean value
refers to the channel from which the trigger signal originates.

The DC mean value is displayed with an algebraic sign (e.g.
”dc:Y1 501mV” resp. ”dc:Y1 -501mV). Overranging is indicated
by ” < ” resp. ” > ” sign (e.g. ”dc:Y1 <1.80V” resp. ”dc:Y1
>1.80V”). Being dependent on a necessary time constant for
mean value creation, the display update requires a few seconds
after a voltage change.
The reading accuracy is dependent on the instrument
specifications (Y deflection tolerance max. 3% from 5 mV/div. to
20 V/div.). Although the tolerances are significantly smaller in

AUTOSET

Summary of Contents for HM504-2

Page 1: ...furbished and purchasing new Financing options such as Financial Rental and Leasing are also available on application TMG will assist if you are unsure whether this model will suit your requirements C...

Page 2: ......

Page 3: ...O s c i l l o s c o p e H M 5 0 4 2 Manual English...

Page 4: ...25 Phase comparison with Lissajous figures 26 Phase difference measurement in DUAL mode Yt 26 Phase difference measurement in DUAL mode 26 Measurement of an amplitude modulation 26 Triggering and tim...

Page 5: ...trol cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the in terior of the oscilloscope is screened by the cabinet direct radiation can...

Page 6: ...signal level Time Base delay provide high X Magnification of any portion of the signal 100 MHz 4 Digit Frequency Counter Cursor and Automatic Measurement Save Recall Memories for Instrument Settings R...

Page 7: ...ontrols Autoset automatic signal related parameter settings Save and Recall 9 instrument parameter settings Readout display of menu parameters cursors and results Autom Measurement Freq Period Vdc Vpp...

Page 8: ...the handle can be set into diffe rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked STOP Attention When changing t...

Page 9: ...nrequires the horizontal or inclined position STOP Do not obstruct the ventilation holes Specifications are valid after a 20 minute warm up period between 15 and 30 degr C Specifications without toler...

Page 10: ...pprox 10 and the real voltage value is 11 higher The gain reduction error can not be defined exactly as the 3 dB bandwidth of the Y amplifiers differs between 50 MHz and 55 MHz When examining square o...

Page 11: ...pe However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 2...

Page 12: ...nt lowerinputsensitivity must be chosen until the vertical signal height is only 3 8 div With a signal amplitude greater than 160 Vpp and the deflection coefficient VOLTS DIV in calibrated condition a...

Page 13: ...eristic resistance of the cable as a rule 50 Ohm Especially when transmitting square and pulse signals a resistor equal to the characteristic impedance of the cable must also be connected across the c...

Page 14: ...d by reducing the READOUT intensity RO The INTENS FOCUS control knob adjusts the READOUT intensity Turning this knob clockwise increases and counter clockwisedecreasestheintensity Onlytheminimumrequir...

Page 15: ...Rot with Int trace rotation by use of INTENS FOCUS knob and allows you to compensate the influence of the Earth s magneticfieldonthetracedeflection Toavoidmisadjustment due to unavoidable deflection...

Page 16: ...selected This allows for a different slope setting for the triggered DELAY DTR time base mode 10 TR Trigger indicator LED The TR LED is lit in Yt mode if the triggering conditions are met for the firs...

Page 17: ...the readout e g Y1 deflection coefficient input coupling The symbolizes calibrated measuring conditions and is replaced by the symbol in uncalibrated conditions VAR The vernier variable function is de...

Page 18: ...e internal trigger signal originating from channel I or channel II or the external trigger signal can be chosen DUAL and Addition mode Briefly pressing opens the trigger source Pulldown menu with the...

Page 19: ...s the way the trigger signal is connected to the trigger unit AC DC content suppressed DC peak value detection inactive HF high pass filter cuts off frequencies below approx 50 kHz trigger point symbo...

Page 20: ...e trace start position which is present after switching over from sea to del This enables lower time deflection coefficient settings for signal expansion del In DELAY mode a trigger event does not sta...

Page 21: ...PUT CH II BNC socket This BNC socket is the signal input for channel II The outer ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of t...

Page 22: ...ifferent frequency responses of the trigger amplifier and Y amplifier cause a reduction of the measurement accuracy If relatively low frequency signals 20 Hz are present the measurement value continuo...

Page 23: ...io X measurement causes the display of two long and one short CURSOR lines and is enabled inYt time base mode only The unit to be displayed must be selected by briefly pressing the UNIT 35 pushbutton...

Page 24: ...es are displayed Briefly pressing SOURCE selects the channel and it s deflection coefficient for the measurement The CURSOR lines must be set to the signal according to the selected channel b DUAL mod...

Page 25: ...ubmenus Main Menue 50 Ohm For high impedance loads Oscilloscope approx 1 M Ohm Digital Voltmeter approx 10 MOhm the output voltage is either 0 2 Volt DC or 0 2 Vpp AC square wave Under First Time Oper...

Page 26: ...n generator provides a square wave signal with selectable frequencies and a very fast rise time 4 ns from the output socket below the CRT screen As the square wave signals are used for probe compensat...

Page 27: ...the calibrator output via a suitable probe The voltage provided by the probe to a high impedance input 1 MOhm II 15 30 pF will correspond to the division ratio of the probe used 10 1 20 mVpp output S...

Page 28: ...the phosphor at too high a brightness setting INTENS setting which causes either a lasting loss of brightness or in the extreme case complete destruction of the phosphor at this point Phase difference...

Page 29: ...a sine signal Except when external trigger is used the trigger threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger...

Page 30: ...formation regarding this item can be noted in the Data Sheet The coupling setting TRIG MODE 20 and indication are described under Controls and Readout As the automatic triggering does not work below 2...

Page 31: ...er threshold less than 0 5 div It is therefore particularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in a circuit In this trigger mode the slope di...

Page 32: ...gnals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use of the holdoff control simplifies t...

Page 33: ...osen in del and dTr mode is stored and automatically set after activating one of those modes If the stored time coefficient in del dTr mode was higher than the actual value in sea search mode the time...

Page 34: ...me base mode I X x10 magnifier switched off I Optimum X and Y position settings I Trace and readout visible If DC trigger coupling had been selected AC trigger coupling will not be chosen and the auto...

Page 35: ...w typical patterns displayed by the various components under test I Open circuit is indicated by a straight horizontal line I Short circuit is shown by a straight vertical line Testing Resistors If th...

Page 36: ...or ground for the same terminal is then absolutely necessary A connection inversion effects a rotation of the test pattern by 180 degrees about the centre point of the scope graticule Pay attention t...

Page 37: ...ation The oscilloscope is supplied with a serial interface for control purposes The interface connector 9 pole D SUB female is located on the rear of the instrument Via this bidirectional port the ins...

Page 38: ...RECALL INPUT CHI X AC DC GND max 400Vp 1M II 18pF INPUT CHI X max 400Vp 1M II 18pF TRIG EXT max 100Vp INP Z x1 x10 AC DC GND x1 x10 50 MHz ANALOG OSCILLOSCOPE HM504 2 Instruments CAT I TRACE ROT CT GL...

Page 39: ...37 Subject to change without notice Front Panel HM504 2...

Page 40: ......

Page 41: ...LEER...

Page 42: ...chwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes Spectrum Analyzer P...

Reviews: