background image

19

Subject to change without notice

The current function is indicated by the VAR LED. The

 TIME/

DIV. 

knob functions as a vernier when the VAR LED is

switched on, but the time base setting remains calibrated
until the (vernier) knob is operated. The readout now indicates
e.g. 

”>10ms”

 instead of ”10ms”. Rotating further

anticlockwise increases the time deflection coefficient
(uncalibrated) until the maximum is reached indicated by a
beep. Rotating the knob clockwise has the opposite effect.
Now, the vernier is again in the calibrated position and the
symbol

  ”>” 

extinguishes.

Underneath the front panel sector described above, the BNC
sockets and two pushbuttons are located.

[25] INPUT CH I (X)

  – BNC socket.

This BNC socket is the signal input for channel I. The outer
(ground) connection is galvanically connected to the
instrument ground and consequently to the safety earth
contact of the line/mains plug. The AC/DC/GND pushbutton

[26]

 is assigned to the input.

In 

XY

 mode, signals at this input are used for the 

X

 deflection.

[26] AC/DC/GND x1/x10 

 – Pushbutton with several functions.

AC/DC/GND:

Briefly pressing this pushbutton opens the input coupling
Pulldown menu if a channel mode is present in which
channel I is activated.
The following input couplings are available: 

AC

DC

 and 

GND

(ground). Please note “B: Menu Display and Operation”.

After the Pulldown menu has extinguished, the READOUT
displays the present input coupling at the bottom right hand
of 

”Y1: deflection coefficient”

; the “~“ symbol indicates

AC, the “=” symbol DC and “GND” is for ground.
The

 GND 

setting disables the input signal and the

 VOLTS/

DIV [14] 

knob. Then in automatic trigger mode (Yt) the

undeflected trace is visible representing the 0 Volt trace
position; in XY mode the X deflection is deactivated.

x1/x10:

Probe factor selection is performed by pressing and holding
the pushbutton. This selects the indicated deflection
coefficient of channel I displayed in the readout, between 1:1
and 10:1. In condition 10:1, the probe factor is thus indicated
by a probe symbol displayed by the readout in front the
channel information (e.g. 

”probe symbol”, Y1...

). In the

case of cursor voltage measurement, the probe factor is
automatically included.

Please note:
The symbol should not be activated unless a x10 (10:1)
attenuator probe is used.

[27] Ground 

socket 4mm banana socket galvanically connected

to safety earth. This socket can be used as reference poten-
tial connection for

 DC 

and low frequency signal measurement

purposes and in

 “Component Tester” 

mode.

[28] INPUT CH 

II  – BNC socket.

This BNC socket is the signal input for channel II. The outer
(ground) connection is galvanically connected to the
instrument ground and consequently to the safety earth
contact of the line/mains plug. The AC/DC/GND pushbutton
(29) is assigned to the input.

In 

XY

 mode, signals at this input are used for the Y deflection.

[29] AC/DC/GND x1/x10

  – Pushbutton with several functions.

AC/DC/GND:

Briefly pressing this pushbutton opens the input coupling
Pulldown menu if a channel mode is present in which
channel II is activated.

The following input couplings are available: 

AC

DC

 and 

GND

(ground). Please note “B: Menu Display and Operation”.

After the Pulldown menu has extinguished, the READOUT
displays the present input coupling at the bottom right hand
of 

”Y2: deflection coefficient”

; the “~“ symbol indicates

AC, the “=” symbol DC and “GND” is for ground.

The

 GND 

setting disables the input signal and the

 VOLTS/

DIV (18) 

knob. Then in automatic trigger mode (Yt) the

undeflected trace is visible representing the 0 Volt trace
position; in XY mode the Y deflection is deactivated.

x1/x10:

Probe factor selection is performed by pressing and holding
the pushbutton. This selects the indicated deflection
coefficient of channel II displayed in the readout, between
1:1 and 10:1. In condition 10:1 the probe factor is thus
indicated by a probe symbol displayed by the readout in front
the channel information (e.g. 

”probe symbol”, Y2...

). In the

case of cursor voltage measurement, the probe factor is
automatically included.

Please note:
The symbol should not be activated unless a x10 (10:1)
attenuator probe is used.

[30] TRIG. EXT. / INPUT (Z)

  – BNC socket with two functions.

The outer (ground) connection is galvanically connected to
the instrument ground and consequently to the safety earth
contact of the line/mains plug. The input impedance is
approx. 1 MOhm II 20pF.
Briefly pressing the Z ON/OFF   VAR 

[24]

 pushbutton switches

over the function of this socket.

TRIG. EXT:

The

 BNC 

socket serves as external trigger signal input, if

external triggering is selected.
The trigger coupling depends on the

 TRIG. MODE [20]

setting.

Z-Input:

If neither

 “Component Tester”, 

delayed time base mode

(“sea”, “del” or “dTr”)

 

nor external trigger coupling

 (“ext”)

is chosen, the socket is operative as a 

(trace intensity

modulation) input.
High TTL level (positive logic) affects blanking, low level
gives unblanking. No higher voltages than +5 Volt are
permitted.
Below the CRT are the controls for the readout, the component
tester and the squarewave calibrator with their outputs.

[31]

Controls and Readout

INPUT CHI (X)

AC/DC/GND

max.
400Vp

1M

 II 

18pF

INPUT CHI (X)

max.
400Vp

1M

 II 

18pF

TRIG. EXT.

max.
100Vp

INP. 
(Z)

x1 / x10

AC/DC/GND

x1 / x10

CAT

 

I

!

!

!

26

27

28

29

30

25

Summary of Contents for HM504-2

Page 1: ...furbished and purchasing new Financing options such as Financial Rental and Leasing are also available on application TMG will assist if you are unsure whether this model will suit your requirements C...

Page 2: ......

Page 3: ...O s c i l l o s c o p e H M 5 0 4 2 Manual English...

Page 4: ...25 Phase comparison with Lissajous figures 26 Phase difference measurement in DUAL mode Yt 26 Phase difference measurement in DUAL mode 26 Measurement of an amplitude modulation 26 Triggering and tim...

Page 5: ...trol cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the in terior of the oscilloscope is screened by the cabinet direct radiation can...

Page 6: ...signal level Time Base delay provide high X Magnification of any portion of the signal 100 MHz 4 Digit Frequency Counter Cursor and Automatic Measurement Save Recall Memories for Instrument Settings R...

Page 7: ...ontrols Autoset automatic signal related parameter settings Save and Recall 9 instrument parameter settings Readout display of menu parameters cursors and results Autom Measurement Freq Period Vdc Vpp...

Page 8: ...the handle can be set into diffe rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked STOP Attention When changing t...

Page 9: ...nrequires the horizontal or inclined position STOP Do not obstruct the ventilation holes Specifications are valid after a 20 minute warm up period between 15 and 30 degr C Specifications without toler...

Page 10: ...pprox 10 and the real voltage value is 11 higher The gain reduction error can not be defined exactly as the 3 dB bandwidth of the Y amplifiers differs between 50 MHz and 55 MHz When examining square o...

Page 11: ...pe However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 2...

Page 12: ...nt lowerinputsensitivity must be chosen until the vertical signal height is only 3 8 div With a signal amplitude greater than 160 Vpp and the deflection coefficient VOLTS DIV in calibrated condition a...

Page 13: ...eristic resistance of the cable as a rule 50 Ohm Especially when transmitting square and pulse signals a resistor equal to the characteristic impedance of the cable must also be connected across the c...

Page 14: ...d by reducing the READOUT intensity RO The INTENS FOCUS control knob adjusts the READOUT intensity Turning this knob clockwise increases and counter clockwisedecreasestheintensity Onlytheminimumrequir...

Page 15: ...Rot with Int trace rotation by use of INTENS FOCUS knob and allows you to compensate the influence of the Earth s magneticfieldonthetracedeflection Toavoidmisadjustment due to unavoidable deflection...

Page 16: ...selected This allows for a different slope setting for the triggered DELAY DTR time base mode 10 TR Trigger indicator LED The TR LED is lit in Yt mode if the triggering conditions are met for the firs...

Page 17: ...the readout e g Y1 deflection coefficient input coupling The symbolizes calibrated measuring conditions and is replaced by the symbol in uncalibrated conditions VAR The vernier variable function is de...

Page 18: ...e internal trigger signal originating from channel I or channel II or the external trigger signal can be chosen DUAL and Addition mode Briefly pressing opens the trigger source Pulldown menu with the...

Page 19: ...s the way the trigger signal is connected to the trigger unit AC DC content suppressed DC peak value detection inactive HF high pass filter cuts off frequencies below approx 50 kHz trigger point symbo...

Page 20: ...e trace start position which is present after switching over from sea to del This enables lower time deflection coefficient settings for signal expansion del In DELAY mode a trigger event does not sta...

Page 21: ...PUT CH II BNC socket This BNC socket is the signal input for channel II The outer ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of t...

Page 22: ...ifferent frequency responses of the trigger amplifier and Y amplifier cause a reduction of the measurement accuracy If relatively low frequency signals 20 Hz are present the measurement value continuo...

Page 23: ...io X measurement causes the display of two long and one short CURSOR lines and is enabled inYt time base mode only The unit to be displayed must be selected by briefly pressing the UNIT 35 pushbutton...

Page 24: ...es are displayed Briefly pressing SOURCE selects the channel and it s deflection coefficient for the measurement The CURSOR lines must be set to the signal according to the selected channel b DUAL mod...

Page 25: ...ubmenus Main Menue 50 Ohm For high impedance loads Oscilloscope approx 1 M Ohm Digital Voltmeter approx 10 MOhm the output voltage is either 0 2 Volt DC or 0 2 Vpp AC square wave Under First Time Oper...

Page 26: ...n generator provides a square wave signal with selectable frequencies and a very fast rise time 4 ns from the output socket below the CRT screen As the square wave signals are used for probe compensat...

Page 27: ...the calibrator output via a suitable probe The voltage provided by the probe to a high impedance input 1 MOhm II 15 30 pF will correspond to the division ratio of the probe used 10 1 20 mVpp output S...

Page 28: ...the phosphor at too high a brightness setting INTENS setting which causes either a lasting loss of brightness or in the extreme case complete destruction of the phosphor at this point Phase difference...

Page 29: ...a sine signal Except when external trigger is used the trigger threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger...

Page 30: ...formation regarding this item can be noted in the Data Sheet The coupling setting TRIG MODE 20 and indication are described under Controls and Readout As the automatic triggering does not work below 2...

Page 31: ...er threshold less than 0 5 div It is therefore particularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in a circuit In this trigger mode the slope di...

Page 32: ...gnals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use of the holdoff control simplifies t...

Page 33: ...osen in del and dTr mode is stored and automatically set after activating one of those modes If the stored time coefficient in del dTr mode was higher than the actual value in sea search mode the time...

Page 34: ...me base mode I X x10 magnifier switched off I Optimum X and Y position settings I Trace and readout visible If DC trigger coupling had been selected AC trigger coupling will not be chosen and the auto...

Page 35: ...w typical patterns displayed by the various components under test I Open circuit is indicated by a straight horizontal line I Short circuit is shown by a straight vertical line Testing Resistors If th...

Page 36: ...or ground for the same terminal is then absolutely necessary A connection inversion effects a rotation of the test pattern by 180 degrees about the centre point of the scope graticule Pay attention t...

Page 37: ...ation The oscilloscope is supplied with a serial interface for control purposes The interface connector 9 pole D SUB female is located on the rear of the instrument Via this bidirectional port the ins...

Page 38: ...RECALL INPUT CHI X AC DC GND max 400Vp 1M II 18pF INPUT CHI X max 400Vp 1M II 18pF TRIG EXT max 100Vp INP Z x1 x10 AC DC GND x1 x10 50 MHz ANALOG OSCILLOSCOPE HM504 2 Instruments CAT I TRACE ROT CT GL...

Page 39: ...37 Subject to change without notice Front Panel HM504 2...

Page 40: ......

Page 41: ...LEER...

Page 42: ...chwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes Spectrum Analyzer P...

Reviews: