background image

Subject to change without notice

28

met (

please note “Controls and Readout”

). In the case of

chopped DUAL mode, selecting alternate trigger mode
automatically sets the instrument to alternate DUAL mode. The
trigger point symbol and the peak value detection (in automatic
trigger mode) are internally deactivated. Only the following trigger
coupling modes are available in alternate trigger mode: AC, DC,
NR, HF and LF.

With alternate triggering it is possible to trigger two signals which
are different in frequency (asynchronous). Each input signal must
be of sufficient height to enable trigger. To avoid trigger problems
due to different DC voltage components, AC input coupling for
both channels is recommended.

The internal trigger source is switched in alternate trigger mode
in the same way as the channel switching system in DUAL
alternate mode, i.e. after each time base sweep. Phase difference
measurement is not possible in this trigger mode as the trigger
level and slope setting are equal for both signals. Even with 180°
phase difference between both signals, they appear with the
same slope direction. If signals are applied with a high frequency
ratio (difference), the trace intensity then becomes reduced if
the time base is set to smaller time coefficients (faster sweep).
This happens as the number of sweeps does not increase
because it depends on the lower frequency signal, but with a
faster sweep the phosphor becomes less activated.

External triggering

The external trigger input is activated with the aid of the

 TRIG.

(20) 

pushbutton (

see “Controls and Readout”

), if the trigger

coupling is not set to line/mains trigger coupling. Then the internal
trigger source is deactivated.

As the external trigger signal applied at the TRIG. EXT socket
normally has no relation to the signal height of the displayed
signal, the trigger point symbol is switched off. The external
trigger voltage must have a minimum amplitude of 0.3V

pp

 and

should not increase above 3V

pp

. The input impedance of the

TRIG. EXT. socket is approx. 1M

 II 20pF.

The maximum input voltage of the input circuit is 100V
(DC+peak AC).

The external trigger voltage may have a completely different form
from the test signal voltage, but must be synchronous with the
test signal. Triggering is even possible in certain limits  with whole
number multiples or fractions of the test frequency.

It must be noted that a different phase angle between the
measuring and the triggering signal may cause a display not
coinciding with the slope selection setting.

The trigger coupling selection can also be used in external
triggering mode.

Trigger indicator “TR”

The following description applies to the

 “TR” LED

.

 Please note

item (14) under “Controls and Readout”

.

An LED on condition indicates that the trigger signal has a
sufficient amplitude and the trigger level control setting is correct.
This is valid with automatic and with normal triggering. By
observing the trigger LED, sensitive trigger level adjustment is
possible when normal triggering is used, particularly at very low

signal frequencies. The indication pulses are of only 100ms
duration.

Thus for fast signals the LED appears to glow continuously, for
low repetition rate signals, the LED flashes at the repetition rate
or at a display of several signal periods  not only at the start of
the sweep at the left screen edge, but also at each signal period.
In automatic triggering mode the sweep generator starts
repeatedly without test signal or external trigger voltage. If the
trigger signal frequency decreases the trigger automatic
repetition frequency the sweep generator starts without awaiting
the trigger pulse. This causes an untriggered display and a flashing
trigger LED.

HOLD OFF-time adjustment

For instrument specific information

 please note DEL.POS. - HO

(24) in section “Controls and Readout”

.

If it is found that a trigger point cannot be found on extremely
complex signals, even after careful adjustment of the trigger level
control, a stable display may often be obtained using the holdoff
control. This facility varies the holdoff time between two sweep
periods approx. up to the ratio 10:1. Pulses or other signal
waveforms appearing during this off period cannot trigger the
time base. Particularly with burst signals or aperiodic pulse trains
of the same amplitude, the start of the sweep can be delayed
until the optimum or required time.

A very noisy signal or a signal with a higher interfering frequency
is at times displayed double. It is possible that trigger level
adjustment only controls the mutual phase shift, but not the
double display. The stable single display of the signal, required
for evaluation, is easily obtainable by expanding the hold off time
until one signal is displayed.

A double display is possible with certain pulse signals, where
the pulses alternately show a small difference of the peak

Triggering and Time Base

Fig. 1 shows a case where the holdoff control is in the minimum

position and various different waveforms are overlapped on

the screen, making the signal observation unsuccessful.

Fig. 2 shows a case where only the desired parts of the signal are

stably displayed.

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: