background image

29

Subject to change without notice

amplitudes. Only a very exact trigger level adjustment makes a
single display possible. The use of the holdoff control simplifies
the right adjustment.

After specific use the holdoff control should be reset into its
calibration detent (fully ccw), otherwise the brightness of the
display is reduced drastically. The function is shown in the  figures
1 and 2.

B-Time Base (2nd Time Base)/ Triggering after Delay

Please note instrument specific information in section “Controls
and readout”. The most important controls regarding this mode
are 

A/ALT - B (26), DEL.TRIG. - VAR. (27), TIME/DIV. (25), DEL.

POS.- HO (24) and TRS (8).

As mentioned before, triggering starts the time base sweep and
unblanks the beam. After the maximum X deflection to the right,
the beam is blanked and flies back to the (left) start position.
After the hold off time has elapsed the sweep is started
automatically by the automatic trigger or the next trigger signal.

In normal triggering mode the automatic trigger is switched off
and the sweep will only start on receipt of a trigger signal. As the
trigger point is always at the trace start position, trace expansion
in X direction with the aid of the time base is limited to the display
always from the trace start.

Parts of the signal to be expanded which are displayed near the
trace end (right side of the screen) are lost when the time base
speed is increased (time coefficient reduced). The X expansion
by 10 may also be used, but again the expansion factor is limited.
The B time base solves such problems.

First the A time coefficient should be in a suitable setting to display
a complete signal period. After setting the instrument to alternate
time base mode (A and B), the signal part to be expanded must
be selected with the aid of the intensified sector on the A time
base trace. The B time base deflection coefficient should be at
least a tenth of the A time coefficient.

Higher A/B ratios (e.g. A: 1ms/div, B: 1µs/div = 1000fold
expansion) result in a drastically reduced B trace intensity. Within
certain limits this can be compensated by the intensity setting.
To avoid high contrast between the A and B time base trace,
change from alternate (A and B) to B time base mode. Then the
A trace is invisible.
The start position of the intensified sector on the A trace can be
stated as the delay time. It is related to the A trace start and the
A time coefficient.

The delay function delays the B trace start by a variable time
from the A trigger point. This allows the sweep to begin on any
portion of a signal (except the first millimeters). If the display
shows jitter, it is possible to select triggering of the B time base
after the elapsed delay time.

AUTOSET

The instrument specific information regarding this function is part
of the section

 “Controls and Readout” paragraph AUTOSET

(2)

. As also mentioned in that section all controls are electronically

selected with the exception of the

 POWER

 and calibrator

frequency (CAL 1kHz/1MHz) pushbuttons as well as the

 FOCUS

and

 TR 

(trace rotation) control. Thus automatic signal related

instrument set up in Yt (time base) mode is possible. In most
cases no additional manual instrument setting is required.

Briefly pressing the

 AUTOSET 

pushbutton causes the instrument

to switch over to the last Yt mode settings regarding CH I, CH II
and DUAL. If the instrument was operated in Yt mode, the actual
setting will not be affected with the exception of ADD mode
which will be switched off. At the same time the attenuator(s)
(VOLTS/DIV) are automatically set for a signal display height of
approx. 6 div in single channel mode or if in DUAL mode for
approx. 4 div height for each channel.

This and the following explanation regarding the automatic time
coefficient setting assumes that the pulse duty factor of the input
signal is approx. 1:1. The time deflection coefficient is also set
automatically for a display of approx. 2 signal periods. The time
base setting occurs randomly if complex signals consisting
several frequencies e.g. video signals are present. If cursor
voltage measurement is selected, AUTOSET also affects the
position of the CURSOR lines. Please note 

AUTOSET (2)

 in

section ”Controls and Readout”.

AUTOSET 

sets the instrument automatically to the following

operating conditions:

AC or DC input coupling unaltered

or in GD condition the previously used setting

Internal triggering (channel I or channel II)

Automatic peak (value) triggering

Trigger level in electrical midrange position

Y deflection coefficient(s) calibrated

A time base deflection coefficient calibrated

AC trigger coupling (except DC trigger coupling)

Alternate or B time base mode switched off

X x10 magnifier switched off

Automatic X and Y position settings

A visible trace and readout intensities.

If DC trigger coupling had been selected, AC trigger coupling will
not be chosen and the automatic trigger is operative without the
peak value detection.
The X position is set to the CRT center as well as the Y-position in
CH I or CH II mode.

 Small deviations due to tolerances of the

CRT and other components are unavoidable. 

In DUAL mode

the channel I trace is set to the upper half and the channel II
trace to the lower half of the CRT.

The 1mV/div and 2mV/div deflection coefficients will not be
selected by AUTOSET as the bandwidth is reduced on these
settings.

Attention!
If a signal is applied with a pulse duty factor of approx.
400:1 or larger, an automatic signal display cannot be
performed. The pulse duty factor causes too low a Y
deflection coefficient (sensitivity too high) and too high a
time deflection coefficient (time base speed too slow) and
results in a display in which only the baseline is visible.

AUTOSET

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: