background image

Subject to change without notice

6

20°

10°

A

E

D

C

B

Important hints

This oscilloscope is easy to operate. The logical arrangement of
the controls allows anyone to quickly become familiar with the
operation of the instrument, however, experienced users are also
advised to read through these instructions so that all functions
are understood.
Immediately after unpacking, the instrument should be checked
for mechanical damage and loose parts in the interior. If there is
transport damage, the supplier must be informed immediately.
The instrument must then not be put into operation.

Symbols

ATTENTION - refer to manual

Danger - High voltage

Protective ground (earth) terminal

Use of tilt handle

To view the screen from the best angle, there are three different
positions (C, D, E) for setting up the instrument. If the instrument
is set down on the floor after being carried, the handle
automatically remains in the upright carrying position (A). In order
to place the instrument onto a horizontal surface, the handle
should be turned to the upper side of the oscilloscope (C). For
the D position (10° inclination), the handle should be turned to
the opposite direction of the carrying position until it locks in
place automatically underneath the instrument.

For the E position (20° inclination), the handle should be pulled
to release it from the D position and swing backwards until it
locks once more. The handle may also be set to a position for
horizontal carrying by turning it to the upper side to lock in the B
position. At the same time, the instrument must be lifted, because
otherwise the handle will jump back.

Safety

This instrument has been designed and tested in accordance
with IEC Publication 1010-1, Safety requirements for electrical
equipment for measurement, control, and laboratory use. The
CENELEC regulations EN 61010-1 correspond to this standard. It
hasleft the factory in a safe condition. This instruction manual
contains important information and warnings which have to be
followed by the user to ensure safe operation and to retain the
oscilloscope in a safe condition.

The case, chassis and all measuring terminals are connected to
the protective earth contact of the appliance inlet. The instrument
operates according to Safety Class I (three conductor power cord
with protective earthing conductor and a plug with earthing
contact). The accessible metal parts (case, sockets, jacks) and
the mains/line supply contacts (live, neutral) have been tested
against insulation breakdown with 2200V DC. The mains/line plug
shall only be inserted in a socket outlet provided with a protective
earth contact. The protective action must not be negated by the
use of an extension cord without a protective conductor.

Most cathode-ray tubes develop X-rays. However, the dose
equivalent rate falls far below the maximum permissible value of
36pA/kg (0.5mR/h).

Whenever it is likely that protection has been impaired, the
instrument shall be made inoperative and be secured against any
unintended operation. The protection is likely to be impaired if, for
example, the instrument

shows visible damage,

fails to perform the intended measurements,

has been subjected to prolonged storage under unfavorable
conditions (e.g. in the open or in moist environments),

has been subject to severe transport stress (e.g. in poor
packaging).

Intended purpose and operating conditions

Attention!
This instrument must be used only by qualified experts who
are aware of the risks of electrical measurement.

Due to safety reasons the instrument must only be connected to
a properly installed power outlet, containing a protective earth
conductor. The protective earth connection must not be broken.
The mains/line plug must be inserted before connections are
made to measuring circuits.

CAT I

This instrument is intended for measurement of circuits either
not at all or not directly connected to mains/line (CAT I). Direct
measuring (without galvanic separation) within measuring
category II, III and IV circuits is inadmissible.
Circuits of a device under test are not connected directly with
mains/line if the device under test is operated via an isolation
transformer (Safety Class II). It is also possible to use suitable
transducers (e.g. Current Probe) that meet the demands of
protective class II, to measure indirect at mains/line. The
transducer’s measuring category must be observed.

Measuring Categories

The measuring category relates to transients on mains/line.
Transients are short and fast voltage and current changes that
may appear periodically or aperiodically. The amplitude of
transients increases with decreasing distance to the low voltage
installation source.

Category IV:

 has the highest demands on isolation etc. and is

required for measurement at the low voltage installation source
(e.g. at the supply meter).

Category III:

 is for measurement within the building installation

at e.g. terminal block, power switch, fixed motor, wall outlet etc.

Category II:

 measurement at circuits connected to the low

voltage installation such as home appliances, portable tools etc.

Important hints

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: