background image

17

Subject to change without notice

20ms/div up to 20ns/div (without X-MAG. x10) but the
availability depends on the A time base setting. The internal
control of the oscilloscope prevents the B time deflection
coefficient from becoming higher than the A deflection
coefficient, as such an operation condition would make no
sense.

If the A time base setting is 200µs/div the B time base range
from 20ms/div up to 500µs/div is not available and the
maximum time deflection coefficient for B would be 200µs/
div. In the last named condition the change of the A time
base from 200µs/div to 100µs/div switches the B time base
also to 100µs/div. However the B time base setting remains
unchanged if the A time base is set to 500µs/div.

As already mentioned under

 DUAL (19) 

the channel switching

depends on the time deflection coefficient setting. In the
time base ranges from 500ms/div to 500µs/div chopped
(CHP) channel switching is automatically selected, through
which the switching occurs constantly during the time base
sweeps. Alternate (ALT) channel switching is automatically
used in all other time deflection coefficient settings (200µs/
div - 20ns/div). In the latter case the active channel is switched
off and the previously inactive channel is switched on after
the completion of each time base sweep. To avoid
interference in chopped mode, or to make both channels
appear simultaneously visible, the actual setting (

ALT 

or

 CHP

)

displayed in the readout can be overwritten and changed to
the opposite mode. This is carried out by simultaneously
pressing and holding the

 CHI (18) 

and the

 DUAL (19)

pushbuttons.

(26) A/ALT – B

Pushbutton for time base mode selection.

The instrument contains two time bases designated A and
B. With the aid of the B time base, signal parts displayed by
the A time base can be expanded in X-direction. The
expansion ratio depends on the time deflection coefficient
ratio of both time bases (e.g. “A: 100µs”, “B:1µs” = 100).
With higher expansion ratios the B time base trace intensity
reduces.

Each time the

 A/ALT 

pushbutton is briefly pressed, the time

base mode changes in the sequence A - alternate A and B -
A. The actual setting is displayed in the readout.

A:

In A time base mode the

 TIME/DIV.  (25) 

control knob is

operative only for this time base. The readout then only
displays the A time coefficient. The time base settings for
this condition are stored if the time base mode is changed.

ALT:

If alternate (A and B = ALT) time base mode is selected, the

TIME/DIV (25) 

knob only controls the B time base switch or

vernier function. The alternate time base mode is a
subfunction of the B time base mode where both time base
traces are displayed. Consequently the readout displays both
time deflection coefficients (e.g. “A:100µs B:1µs”). Unlike
the former A time base mode, an intensified sector is also
visible on the A trace. This sector indicates the signal part
which is displayed by the B time base.
The intensified segment can be shifted horizontally by the

DEL. POS. (24) 

control knob continuously (if the B time base

is operated in free run conditions). The difference between

the start of the A time base trace and the beginning of the
intensified sector shows the delay time. This information is
also displayed in the readout as an approximative value (e.g.

t:2.5ms”) related on the calibrated A time coefficient

(uncalibrated e.g. “

t>2.5ms”). The width of the segment

decreases when the B time coefficient is set to a lower value
(higher time deflection speed). For better reading, the vertical
position of the B trace position can be shifted (

please note

“TRS” (8)

).

Alternate time base mode causes the display to alter
between A and B time base after each sweep. In alternate
DUAL mode the sequence is channel I with A time base,
channel I with B time base, channel II with A time base and
channel II with B time base.
In alternate time base mode (A alternate B), the

 X-MAG.

x10 (16)

 function only effects the B time base and similarly

the readout.

B:

Pressing and holding this button switches over from A or
alternate time base mode to B time base mode. If B time
base mode is selected, pressing and holding selects alternate
time base mode. Briefly pressing this button switches over
from B time base mode to A time base mode.
In B time base mode the display of the A trace(s), the
intensified sector(s) and the A time coefficient display in the
readout are disabled. As the trace separation

 (TRS) 

is no

longer required under this circumstances, this function is
switched off too. Consequently only the B time coefficient
is displayed by the readout.
When selecting alternate or B time base mode, the triggered
or free run state depends on the last chosen setting.

(27) DEL.TRIG. - VAR. 

– Pushbutton with two functions.

DEL. TRIG.:

Each time the pushbutton is briefly pressed, the instrument
switches between free run (untriggered) and triggered B time
base, if alternate (A and B) or B time base mode is present.

The actual setting is displayed in the readout. Instead of the
approximate delay time (

“Dt:...”

) in free run mode, the

readout displays

 “DTr: slope, DC (trigger coupling)” 

in

triggered delay mode. In this mode, the former A time base
trigger settings regarding the trigger mode (automatic or
normal, -coupling, -slope and –level) are stored but still active.
W ith the activated delay trigger the instrument is
automatically set to normal trigger mode and DC trigger
coupling for the B time base. As the instrument contains a
separate trigger unit for the B time base, the trigger level

Controls and Readout

MADE IN GERMANY

A/ALT.

DEL.TRIG

TRS

PUSH
BOTH

PUSH
LONG

VAR.

B

CH I

CH II

DEL.POS.

ADD

CHP.

VAR.

VAR.

DUAL

X-Y

VOLTS / DIV.

VOLTS / DIV.

TIME / DIV.

5V 

       1mV

5V 

       1mV

0.5s 

       20ns

TRIG. MODE

TRIG.

ALT

CHI

AC
DC

HF

NR

LF

TVL
TVF

CHII
EXT

Y-POS. I

Y-POS. II

LEVEL

X-POS.

NM

AT

X-

MAG.

63

-K

304-0040/

00

55

BWL

BW

LIMIT

VAR

VAR

TR

NM

HO

VAR

x10

 

17

18

19

20

22

23

24

25

26

27

21

9

10

12

11

14

16

15

13

8

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: