background image

Subject to change without notice

10

required period T = 7x100x10-9 = 0.7µs
required rec. freq. F = 1:(0.7x10-6) = 1.428MHz.

Signal period T = 1s,
set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.

Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10-3) = 100Hz.

TV-Line frequency F = 15625Hz,
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15 625x10-5) = 6.4div.

Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x103) = 0.25ms/div,
min. time coefficient Tc = 1:(10x103) = 0.1ms/div,
set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(103x0.2x10-3) = 5div.

Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) button: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10-6) = 25MHz,
required period T = 1:(25x106) = 40ns.

If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(X-MAG. (x10) active). In this case, the time interval of interest
can be shifted to the screen center using the X-POS. control.

When investigating pulse or square waveforms, the critical feature
is the rise time of the voltage step. To ensure that transients,
ramp-offs, and bandwidth limits do not unduly influence the
measuring accuracy, the rise time is generally measured between
10% and 90% of the vertical pulse height. For measurement,
adjust the Y deflection coefficient using its variable function
(uncalibrated) together with the Y-POS. control so that the pulse
height is precisely aligned with the 0% and 100% lines of the
internal graticule. The 10% and 90% points of the signal will now
coincide with the 10% and 90% graticule lines. The rise time is
given by the product of the horizontal distance in div between
these two coincident points and the calibrated time coefficient
setting. The fall time of a pulse can also be measured by using
this method.

With a time coefficient of 2ns/div (X x10 magnification active),
the example shown in the above figure results in a total measured
rise time of

t

tot

 = 1.6div x 2ns/div = 

3.2ns

When very fast rise times are being measured, the rise times of
the oscilloscope amplifier and of the attenuator probe has to be
deducted from the measured time value. The rise time of the
signal can be calculated using the following formula.

t

r

 = 

 t

tot

2

 - t

osc

2

 - t

p

2

In this t

tot

 is the total measured rise time, t

osc

 is the rise time of

the oscilloscope amplifier (HM2005 approx. 1.75ns), and t

p

 the

rise time of the probe (e.g. = 1.4ns). If t

tot

 is greater than 16ns,

then t

tot

 can be taken as the rise time of the pulse, and calculation

is unnecessary.

Calculation of the example in the figure above results in a signal
rise time

t

r

 = 

 3.2

2

 – 1.75

2

 – 1.4

= 2.28ns

The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to measure
in any display position and at any signal amplitude. It is only
important that the full height of the signal edge of interest is
visible in its full length at not too great steepness and that the
horizontal distance at 10% and 90% of the amplitude is measured.
If the edge shows rounding or overshooting, the 100% should
not be related to the peak values but to the mean pulse heights.
Breaks or peaks (glitches) next to the edge are also not taken
into account. With very severe transient distortions, the rise and
fall time measurement has little meaning. For amplifiers with
approximately constant group delay (therefore good pulse
transmission performance) the following numerical relationship
between rise time

 tr (in ns)

 and bandwidth 

B (in MHz)

 applies:

Connection of Test Signal

In most cases briefly depressing the

 AUTOSET 

causes a useful

signal related instrument setting. The following explanations refer
to special applications and/or signals, demanding a manual
instrument setting. 

The description of the controls is explained

in the section “Controls and Readout”

.

Caution:
When connecting unknown signals to the oscilloscope
input, always use automatic triggering and set the input
coupling switch to AC (readout). The attenuator should
initially be set to 5V/div.

Sometimes the trace will disappear after an input signal has been
applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is only
3-8div. With a signal amplitude greater than 40Vpp and the
deflection coefficient 

(VOLTS/DIV.)

 in calibrated condition, an

attenuator probe must be inserted before the vertical input. If,
after applying the signal, the trace is nearly blanked, the period
of the signal is probably substantially longer than the set time
deflection coefficient (

TIME/DIV.

). It should be switched to an

adequately larger time coefficient.

The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as 

HZ32

or

 HZ34

, or reduced through a x10 or x100 attenuator probe. The

use of test cables with high impedance circuits is only
recommended for relatively low frequencies (up to approx. 50kHz).

Basics of signal voltage

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: