background image

Subject to change without notice

12

Even if alternating time base mode or B time base mode
was active before, the instrument is switched automatically
to A time base mode.

 Please note “AUTOSET”

.

Automatic CURSOR supported voltage measurement. If
CURSOR voltage measurement is present, the CURSOR
lines are automatically set to the positive and negative peak
value of the signal. The accuracy of this function decreases
with higher frequencies and is also influenced by the signal‘s
pulse duty factor.

In DUAL mode the CURSOR lines are related to the signal
which is used for internal triggering. If the signal height is
insufficient, the CURSOR lines do not change.

(3) RM

The remote control mode can be switched on or off via the
RS232 interface. In the latter case the

 “RM” LED 

is lit and

the electronically selectable controls on front panel are
inactive. This state can be left by depressing the 

AUTOSET

pushbutton provided it was not deactivated via the interface.

(4) INTENS - READOUT

Knob with associated pushbutton and LEDs.
This control knob is for adjusting the A and B traces and
readout intensity. Turning this knob clockwise increases and
turning it counterclockwise decreases the intensity.

The

 READOUT 

pushbutton below is for selecting the function

in two ways.
Depending on the actual time base mode and with the
readout

 (RO) 

not switched off, briefly pressing the 

READOUT

pushbutton switches over the

 INTENS 

knob function

indicated by a LED in the sequences:

A - RO - A in condition A time base,
A - RO - B - A if alternate time base mode is present and
B - RO - B in condition B time base.
XY mode:

A - RO - A.

Component Test: A - RO - A.

Pressing and holding the

 READOUT 

pushbutton switches

the readout on or off. In readout off condition the

 INTENS

knob function can consequently not be set to

 RO

. Briefly

pressing the pushbutton causes the following sequences:

condition

sequence

A time base

A - A

Alternate A/B

A - B - A

B time base

B - B

XY mode

A - A

Component Test

A  -  A

Switching the readout off, may be required if interference is
visible on the signal(s). Such interference may also originate
from the chopper generator if the instrument is operated in
chopped

 DUAL 

mode. All INTENS-settings are stored after

the instrument is switched off.
The

 AUTOSET 

function switches the readout on and selects

A time base mode (A-LED lit). The

 INTENS 

setting for each

function is automatically set to the mean value, if less
intensity was previously selected.

(5) TR

The trace rotation control can be adjusted with a small
screwdriver (

please note “trace rotation TR”

)

(6) FOCUS

This control knob effects both the trace and the readout
sharpness.

(7) SAVE / RECALL

The instrument contains 9 non volatile memories. These can
be used by the operator to save instrument settings and to
recall them. This relates to all controls which are electronically
selected.

Press the

 SAVE 

pushbutton briefly to start the save

procedure. The readout then indicates the letter

 “S”

 followed

by a cipher between 1 and 9, indicating the memory location.
If the instrument settings stored in this memory location
must not be overwritten, briefly press the

 SAVE 

or the

RECALL 

pushbutton to select another memory location. Each

time the

 SAVE 

pushbutton is briefly pressed the memory

location cipher increases until the location number 9 is
reached. The

 RECALL

 pushbutton function is similar but

decreases the memory location cipher until 1 is reached.
Press and hold

 SAVE 

for approx. 3 seconds to write the

instruments settings in the memory and to switch the
associated readout information (e.g. “S8”) off.

To recall a front panel setup, start that procedure by briefly
pressing the

 RECALL 

pushbutton. The readout then indicates

the letter

 “R” 

and the memory location number. If required,

select a different memory location as described above. Recall
the settings by pressing and holding the RECALL pushbutton
for approx. 3 seconds.

If the

 SAVE 

or the

 RECALL 

pushbutton was depressed

inadvertently, briefly press both pushbuttons at the same
time or wait approx. 10 seconds without pressing either
pushbutton to exit that function.

Switching the instrument off results in an automatic SAVE
procedure of the present settings in memory location 9 and
overwrites the data in that location. If the instrument settings
in memory location 9 are of importance, RECALL 9 before
switching the instrument off.

Attention:
Make sure that the signal to be displayed is similar to the
one that was present when the settings were stored. If
the signal is different (frequency, amplitude) to the one
during storage then a distorted display may result.

The setting controls and LED’s for the Y amplifiers, modes,
triggering and time bases are located underneath the sector of
the front panel described before (Responding picture see next
page).

(8) TRS 

– Pushbutton wit associated LED.

The instrument contains a trace separation function which
is required in the alternate time base mode to separate the
B time base trace from the A time base in Y direction.

Controls and Readout

AUTOSET

TR

B

RM

READ

OUT

200 MHz

ANALOG OSCILLOSCOPE

HM2005

Instruments

!

POWER

INTENS

EXIT

SAVE

SET

1

FOCUS

9

RECALL

A

RO

2

1

3

4

5

6

7

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: