background image

7

Subject to change without notice

Field of application

The instrument is specified for operation in industry, light industry,
commercial and residential environments.
The instrument has been designed for indoor use. The permissible
ambient temperature range during operation is +10°C (+50°F) ...
+40°C (+104°F). It may occasionally be subjected to temperatures
b10°C (+50°F) and -10°C (+14°F) without degrading its
safety. The permissible ambient temperature range for storage or
transportation is -40°C (-40°F) ... +70°C (+158°F). The maximum
operating altitude is up to 2200m (non-operating 15000m). The
maximum relative humidity is up to 80%.
If condensed water exists in the instrument it should be
acclimatized before switching on. In some cases (e.g. extremely
cold oscilloscope) two hours should be allowed before the
instrument is put into operation. The instrument should be kept
in a clean and dry room and must not be operated in explosive,
corrosive, dusty, or moist environments. The oscilloscope can be
operated in any position, but the convection cooling must not be
impaired. The ventilation holes may not be covered. For continuous
operation the instrument should be used in the horizontal position,
preferably tilted upwards, resting on the tilt handle.
The specifications stating tolerances are only valid if the instrument
has warmed up for 30minutes at an ambient temperature between
+15°C (+59°F) and +30°C (+86°F). Values without tolerances are
typical for an average instrument.

EMC

This instrument conforms to the European standards regarding
the electromagnetic compatibility. The applied standards are:
Generic immunity standard EN50082-2:1995 (for industrial
environment) Generic emission standard EN50081-1:1992 ( for
residential, commercial and light industry environment).
This means that the instrument has been tested to the highest
standards. 

Please note that under the influence of strong

electromagnetic fields, such signals may be superimposed
on the measured signals.

  Under certain conditions this is

unavoidable due to the instrument’s high input sensitivity, high
input impedance and bandwidth. Shielded measuring cables,
shielding and earthing of the device under test may reduce or
eliminate those effects.

Warranty

HAMEG 

warrants to its Customers that the products it

manufactures and sells will be free from defects in materials
and workmanship 

for a period of 2 years

. This  warranty  shall

not apply to any defect, failure or damage caused by improper
use or inadequate maintenance and care.

 HAMEG 

shall not be

obliged to provide service under this warranty to repair damage
resulting from attempts by personnel other than 

HAMEG

representatives to install, repair, service or modify these products.
In order to obtain service under this warranty, Customers must
contact and notify the distributor who has sold the product. Each
instrument is subjected to a quality test with 10 hour burn-in
before leaving the production. Practically all early failures are
detected by this method. In the case of shipments by post, rail
or carrier the original packing must be used. Transport damages
and damage due to gross negligence are not covered by the
guarantee.
In the case of a complaint, a label should be attached to the
housing of the instrument which describes briefly the faults
observed. If at the same time the name and telephone number
(dialing code and telephone or direct number or department
designation) is stated for possible queries, this helps towards
speeding up the processing of guarantee claims.

Maintenance

Various important properties of the oscilloscope should be
carefully checked at certain intervals. Only in this way is it largely
certain that all signals are displayed with the accuracy on which
the technical data are based. Purchase of the 

HAMEG scope

tester HZ 60

, which despite its low price is highly suitable for

tasks of this type, is very much recommended. The exterior of
the oscilloscope should be cleaned regularly with a dusting brush.
Dirt which is difficult to remove on the casing and handle, the
plastic and aluminum parts, can be removed with a moistened
cloth (99% water +1% mild detergent). Spirit or washing benzene
(petroleum ether) can be used to remove greasy dirt. The screen
may be cleaned with water or washing benzene (but not with
spirit (alcohol) or solvents), it must then be wiped with a dry
clean lint-free cloth. Under no circumstances may the cleaning
fluid get into the instrument. The use of other cleaning agents
can attack the plastic and paint surfaces.

Protective Switch Off

This instrument is equipped with a switch mode power supply. It
has both over voltage and overload protection, which will cause
the switch mode supply to limit power consumption to a
minimum. In this case a ticking noise may be heard.

Power supply

The instrument operates on mains/line voltages between 100V

AC

and 240V

AC

. No means of switching to different input voltages

has therefore been provided.
The power input fuse is externally accessible. The fuse holder
and the 3 pole power connector is an integrated unit. The power
input fuse can be exchanged after the rubber connector is
removed. The fuse holder can be released by lever action with
the aid of a screwdriver. The starting point is a slot located on
contact pin side. The fuse can then be pushed out of the mounting
and replaced.
The fuse holder must be pushed in against the spring pressure
and locked. Use of patched fuses or short circuiting of the fuse
holder is not permissible; 

HAMEG

 assumes no liability whats-

oever for any damage caused as a result, and all warranty claims
become null and void.

Fuse type:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).

Time characteristic: time lag.

Attention!
There is a fuse located inside the instrument within
the switch mode power supply:

Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: fast (F).

The operator must not replace this fuse!

Important hints

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: