background image

Subject to change without notice

30

In such cases it is recommended to select normal triggering and
to set the trigger point approx. 0.5div above or below the trace.
If under one of these conditions the trigger indicator LED is lit,
this indicates the presence of a signal. Then both the time
coefficient and Y deflection coefficient should be reduced. Please
note that a reduction in intensity may occur, which could result
in a blank screen when the physical limits are reached.

Mean Value Display

The DC Mean Value of the input voltage is displayed in place of
the CURSOR line measurement, if the CURSOR measurement
function is not activated. MEAN VALUE 

ON

 must activated under

MISCELLANEOUS (submenu of SETUP) and further conditions
met.

The signal to be measured (AC > 20 Hz) must be applied at input
CH I or CH II with its DC content at the input (i.e. DC coupling
required). Yt (time base) mode and internal triggering (trigger
source CH I or CH II; not alternated triggering) must be used. AC
or DC trigger coupling must be selected. If the above conditions
are not met, “DC:?” will be displayed.

The mean value is acquired using the internal trigger signal
amplifiers. With the exception of DUAL mode, the indicated mean
value is automatically related to the active channel (CH I or CH
II), as the channel selection also selects the assigned trigger
amplifier. In DUAL mode one can select between trigger amplifier
CH I or CH II for triggering. Thus the indicated mean value refers
to the channel from which the trigger signal originates.

The DC mean value is displayed with an algebraic sign (e.g. “DC:
501mV” or “DC: -501mV). Overranging is indicated by “ < “ or “
> “ sign (e.g. “DC<1.80V” or “DC>1.80V”). Being dependent on
a necessary time constant for mean value creation, the display
update requires a few seconds after a voltage change.

The reading accuracy is dependent on the instrument
specifications (vertical deflection tolerance max. 3% from 5mV/
div. to 20V/div.). Although the tolerances are significantly smaller
in reality, other deviations such as unavoidable offset voltages
must be taken into account, which may cause a display deviating
from 0 Volt without signal applied at the input.

The display shows the arithmetic (linear) mean value. The DC
content is displayed if DC or DC with superimposed AC voltages
are applied. In the case of square wave voltages, the mean value
depends on the pulse duty factor.

Component Tester

General

The instrument specific information regarding the control and
terminals are part of

 item (42) 

in section

 “Controls and Readout”

.

The instrument has a built-in electronic Component Tester, which
is used for instant display of a test pattern to indicate whether or
not components are faulty. It can be used for quick checks of
semiconductors (e.g. diodes and transistors), resistors,

capacitors, and inductors. Certain tests can also be made to
integrated circuits. All these components can be tested
individually, or in circuit provided that it is unpowered. The test
principle is fascinatingly simple. A built-in generator delivers a
sine voltage, which is applied across the component under test
and a built-in fixed resistor. The sine voltage across the test object
is used for the horizontal deflection, and the voltage drop across
the resistor (i.e. current through test object) is used for vertical
deflection of the oscilloscope. The test pattern shows a current-
voltage characteristic of the test object.

The measurement range of the component tester is limited and
depends on the maximum test voltage and current (please note
data sheet). The impedance of the component under test is limited
to a range from approx. 20

 to 4.7k

. Below and above these

values, the test pattern shows only short-circuit or open-circuit.
For the interpretation of the displayed test pattern, these limits
should always be born in mind. However, most electronic
components can normally be tested without any restriction.

Using the Component Tester

After the component tester is switched on, the vertical
preamplifier and the time base generator are inoperative. A
shortened horizontal trace will be observed. It is not necessary
to disconnect scope input cables unless in-circuit measurements
are to be carried out. For the component connection, two simple
test leads with 4mm Ø banana plugs, and with test prod, alligator
clip or sprung hook, are required.
The test leads are connected as described in section

 “Controls

and Readout

”.

Caution!
Do not test any component in live circuitry. Remove all
grounds, power and signals connected to the component
under test. Set up Component Tester as stated. Connect
test leads across component to be tested. Observe oscil-
loscope display.

IMPORTANT: Only discharged capacitors should be testet!

Test Pattern Displays

The “Test patterns” shows typical patterns displayed by the
various components under test.

Open circuit is indicated by a straight horizontal line.

Short circuit is shown by a straight vertical line.

Testing Resistors

If the test object has a linear ohmic resistance, both deflecting
voltages are in the same phase. The test pattern expected from
a resistor is therefore a sloping straight line. The angle of slope is
determined by the resistance of the resistor under test.

With high values of resistance, the slope will tend towards the
horizontal axis, and with low values, the slope will move towards
the vertical axis. Values of resistance from 20

 to 4.7k

 can be

approximately evaluated. The determination of actual values will
come with experience, or by direct comparison with a component
of a known value.

Testing Capacitors and Inductors

Capacitors and inductors cause a phase difference between
current and voltage, and therefore between the X and Y deflection,

Mean Value Display

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: