background image

Subject to change without notice

26

When using the internal normal triggering mode, it is possible to
trigger at any amplitude point of a signal edge, even with very
complex signal shapes, by adjusting the trigger LEVEL control. If
the signal applied at the Y input is used for triggering (internal
trigger source), its adjusting range is directly dependent on the
display height, which should be at least 0.5div. If it is smaller
than 1div, the trigger

 LEVEL 

adjustment needs to be operated

with a sensitive touch. In the external normal triggering mode,
the same applies to approx. 0.3Vpp external trigger voltage
amplitude.

Other measures for triggering of very complex signals are the
use of the time base variable control and hold off time control,
mentioned later.

 - Slope

Please note item (12) in section “Controls and Readout” for
instrument specific information.

The actual slope setting is displayed in the readout. The setting
is not changed by the

 AUTOSET 

function. The slope setting

can be changed in  alternate- and B time base mode for the B
time base trigger unit if the delay trigger function is active.
The previous slope setting for the A time base trigger is stored
and still active. For further information

 please note “Controls

and Readout”

.

The time base generator can be triggered by a rising or falling
edge of the test signal. Whether the rising or the falling edge is
used for triggering, depends on the slope direction setting. This
is valid with automatic and normal triggering. The positive slope
direction means an edge going from a negative potential and
rising to a positive potential. This has nothing to do with zero or
ground potential and absolute voltage values. The positive slope
may also lie in a negative part of a signal.

However the trigger point may be varied within certain limits on
the chosen edge using the LEVEL control. The slope direction is
always related to the input signal and the non inverted display.

Trigger coupling

Instrument specific information regarding this item can be noted
in the “Data Sheet”. The coupling setting

 ( TRIG. MODE (23)) 

and

indication are described under

 “Controls and Readout”

.

As the automatic triggering does not work below 20Hz, normal
triggering should be used in DC and LF trigger coupling mode.
The coupling mode and accordingly the frequency range of the
trigger signal should meet the signal requirements.

AC:

This is the most frequently used trigger mode. The trigger
threshold increases below and above the frequency limits
mentioned in the data sheet. This filter cuts off both the
DC content of the trigger signal and the lowest frequency
range.

DC:

In this coupling mode the trigger signal is coupled
galvanically to the trigger unit if normal triggering (NM) is
present. Therefore there is no  low frequency limit.

DC triggering is recommended if the signal is to be
triggered with quite slow processes or if pulse signals
with constantly changing pulse duty factors have to be
displayed.

threshold can be stated as vertical display height in div, at which
the time base generator starts, the display is stable, and the
trigger indicator LED lights or flashes. The internal trigger
threshold of the oscilloscope is given as 

 0.5div. When the trigger

voltage is externally supplied, it can be measured in Vpp at that
input. Normally, the trigger threshold may be exceeded up to a
maximum factor of 20.
The instrument has two trigger modes, which are characterized
as Automatic- Peak and Normal-Triggering.

Automatic Peak (Value) -Triggering

Instrument specific information can be drawn from the items

NM - AT - 

 (12) 

and

 LEVEL (13) 

in the section

 “Controls and

Readout”

.

This trigger mode is automatically selected after the

 AUTOSET

pushbutton is pressed. As the peak value detection makes no
sense in combination with

 DC 

and

 TV 

(television) signals, it is

switched off automatically in

 DC

,

 TVL 

and

 TVF 

trigger coupling

conditions as well as in alternate trigger mode. In this case the
automatic is still present, but a wrong trigger level setting causes
an untriggered display.

In automatic trigger mode the sweep generator can run without
a test signal or external trigger voltage. A base line will always
be displayed even with no signal. With an applied AC signal the
peak value triggering enables the user to select the voltage point
on the trigger signal (trigger point), by the adjustment of the
trigger level control. The control range depends on the peak to
peak value of the signal. This trigger mode is therefore called
Automatic Peak (Value)- Triggering. Operation of the scope needs
only correct amplitude and time base settings, for a constantly
visible trace.

Automatic mode is recommended for all uncomplicated
measuring tasks. However, automatic triggering is also the
appropriate operation mode for the “entry” into difficult measuring
problems, e.g. when the test signal is unknown relating to
amplitude, frequency or shape. Presetting of all parameters is
now possible with automatic triggering; the change to normal
triggering can follow thereafter.

The automatic triggering works above 20Hz. The failure of
automatic triggering at frequencies below 20Hz is abrupt.
However, it is not signified by the trigger indicator LED which is
still blinking. Loss of triggering is best recognizable at the left
screen edge (the start of the trace in differing display height).

The automatic peak (value) triggering operates over all variations
or fluctuations of the test signal above 20Hz. However, if the
pulse duty factor of a square-wave signal exceeds a ratio of 100:1,
switching over to normal triggering will be necessary. Automatic
triggering is practicable with internal and external trigger voltage.

Normal Triggering

The description of the most important control elements (

NM -

AT - 

 (12) 

and

 LEVEL (13)

) is part of the section

 “Controls

and Readout”

).

With normal triggering, the sweep can be started by AC signals
within the frequency range defined by the trigger coupling setting.
In the absence of an adequate trigger signal or when the trigger
controls (particularly the trigger LEVEL control) are misadjusted,
no trace is visible.

Triggering and Time Base

Summary of Contents for HM 2005

Page 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Page 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Page 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Page 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Page 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Page 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Page 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Page 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Page 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Page 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Page 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Page 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Page 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Page 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Page 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Page 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Page 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Page 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Page 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Page 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Page 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Page 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Page 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Page 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Page 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Page 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Page 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Page 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Page 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Page 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Page 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Page 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Page 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Page 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Page 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Page 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Reviews: