BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
45
8.0
Connecting Cable and Power
8.1
Cabling
WARNINGS
!
•
Any accessories and cables not approved by Haemonetics
®
used in
conjunction with the BloodTrack
®
Kiosk V5 may increase hazards and
influence compatibility with EMC requirements. Non-approved accessories
and cables must
not
be used.
•
Connections to network points other than those approved by Haemonetics
as part of the system may result in previously unidentified risks to patients,
operators or third parties. These risks should be identified, analyzed,
evaluated and controlled. Contact Haemonetics Technical Support for
guidance
—see
Appendix C
•
The BloodTrack Kiosk V5 must only use a CAT5e/CAT6 shielded Ethernet
cables length 9.8 ft. (3 m) or less to the network point from the BloodTrack
Kiosk V5.
CAUTIONS
•
Care should be taken to insulate all unused cores to prevent shorting.
•
Use
only
Haemonetics-approved peripherals, cabling and connections when
installing the BloodTrack Kiosk V5.
The cabling can be run with the BloodTrack Kiosk V5, locks and UPS mounted in
the final position for the end-users.
The supplied shielded CAT5e network patch lead is connected from the BloodTrack
Kiosk V5 to the network data point provided by the hospital.
The BloodTrack Kiosk V5 requires at least one CAT5e/CAT6 network connection to
function. There is provision for a second CAT5e/CAT6 network connection for
redundancy/backup network connection.
8.2
Power Cord
The mains power supply may run via the optional UPS to the BloodTrack Kiosk V5.
WARNINGS
!
•
Any accessories and cables not approved by Haemonetics used in
conjunction with the BloodTrack Kiosk V5 may increase hazards and