BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
47
Figure 19 Lock connection
—terminal block—pin 1 at top
8.3.3 Serial Port Interface
, the serial port connections are:
Pin
Function
1
12V DC supply
2
Receive data (RXD) to BloodTrack Kiosk V5 from peripheral
(TXD on peripheral)
3
Transmit data (TXD) out from BloodTrack Kiosk V5 to peripheral
(RXD on peripheral)
4
Request to send (RTS)
5
Clear to Send (CTS)
6
0V
Chassis
Shield via conductive P-clip
8.3.4 Relay Output Interface
The BloodTrack Kiosk V5 is available with one, two or three relay outputs for
interfacing to the storage location door lock system on one, two or three separate
storage locations.
Note:
The relay output is only for switching less than or equal to a 48V DC or
48V AC rms signal up to 1A current. If higher current, voltage or mains powered
switching is required, contact Haemonetics Technical Support for advice
—see
Appendix C
Pin 1