Electromagnetic Immunity
The BloodTrack Kiosk V5 is intended for use in the electromagnetic environment
specified below. The customer or operator of the BloodTrack Kiosk V5 should
ensure that it is used in such an environment.
B.2.1 IEC 60601-1-2, Table 201
IEC 60601-1-2:1990, Table 201
Guidance and manufacturer’s declaration―electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The BloodTrack Kiosk V5 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The BloodTrack Kiosk V5 is suitable for use in all
establishments including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
FCC CFR 47 Part 15B: 2010 Class B
Spec Clause
Test Description
Base Standard
15.107
Emissions (AC Power Port)
ANSI 63.4
15.109
Radiated Emissions (Enclosure Port)
ANSI 63.4