English
123
7.
Check that the connection cable runs loosely through the device
and then tighten the cable screw connection firmly again.
8.
The limit-value contacts must be used variably as normally closed
contacts or normally open contacts (each with three connection
terminals with switch symbol).
9.
All externally connected lines must be shielded, all shields are
screwed to the terminal block near to the hinge!
Startup
After the installation of the device, the reagents are placed in the
refrigerator.
Refer to
section Exchange of the reagents
.
Upon full hose connection including drain hose, the power plug can be
plugged in both for the refrigerator and for the measurement device.
This begins the measurement operation immediately. All device settings
are described in the software menu system.
Passcode Protection
To protect the device from unauthorized access or unwanted operation,
a password function can be activated directly at commissioning or later
at any time via a laptop by the HACH-LANGE employee. The passcode
is four digits and each of these consists of the numbers 1–4, which can
be freely programmed.
Shutdown
Note:
Please observe during the handling of the reagents the safety
notes under
section Exchange of the reagents
So that no crystallization or other major impurities impede a subsequent
commissioning, the entire system must be purged for a shutdown of
more than 48 hours with distilled water (
Purge process
1.
Call up [+SERVICE] menu
2.
Unscrew screw caps from the reagent canisters A, C and D
complete with the suction pipes (please unscrew ORIGINAL CAP)
and place in a beaker with distilled water
3.
Pull out the hose from the standard solution, remove screw cap,
unscrew ORIGINAL CAP, remove bottle
4.
Select menu item [PRE PUMPING]
5.
After the rinse process, take the suction pipes out of the beaker
again
6.
Select menu item [PRE PUMPING] again
7.
To finish, release pump cassettes and remove the hoses from the
three hose pinch valves
Analysis part
Dosing hoses are hoses that are routed via the sample- and reagent
pumps. Each dosing hose may only be used on the corresponding
sample- or reagent pump.
Note:
After 3 months, replace the sample dosing hose, the drain hose
and the hose for the standard. Coat the pump rollers of the sample pump
with silicone paste.
After 6 months, replace the dosing hoses for the reagents A, C and D.
N O T I C E
For a long-term shutdown (over 2 days), the valve hoses must be removed from
the pinch valves to provide relief. But because of the return flow risk, the hoses
must first be removed from the respective screw caps (canister lids)! Please
observe during the handling of the reagents the safety notes under
.
N O T I C E
For all work on the dosing- and valve hoses, always remove the dosing hoses
from the respective canisters first. (Return flow hazard). Please observe during
the handling of the reagents the safety notes under
Summary of Contents for PHOSPHAX sigma High Range
Page 5: ...5 II 2 230 V AC 10 50 Hz 145 VA x 540 mm x 600 mm x 400 mm 22 kg...
Page 7: ...7 12 2005 2002 96 Hach Lange Hach Lange...
Page 9: ...9 1 SERVICE 2 3 5 6 3 4 4 3 CEE 10 A STO P 1 2 3 4 5 6 7 5 6...
Page 10: ...10 4 1 2 m SIGMATAX 2 PG LZP 573 5 2 m 5 PG 6 9 mm 1 4 m T 2 5 A L 250 V...
Page 11: ...11 6 SIGMATAX 2 SIGMATAX 2 1 2...
Page 15: ...15 15 PHOSPHAX sigma 1 2 m 5 100 C SERVICE C D 9 48...
Page 16: ...16 3 A C D 6 6 1 2 a SERVICE 18 SERVICE 3 6 Hach Lange 10 30...
Page 17: ...17 Q1 Q2 QA A A PHOSPHAX sigma 11 10 C 5 6 48 6 7 PHOSPHAX...
Page 29: ...29 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Page 32: ...32...
Page 114: ...Deutsch 114...
Page 226: ...Fran ais 226...
Page 282: ...Italiano 282...
Page 370: ...Rom n 370...
Page 452: ...T rk e 452...
Page 456: ...456 2005 8 12 2002 96 EC Hach Lange Hach Lange...
Page 463: ...463 100 C SERVICE C D 9 3 A C D 6 6 1 2 48 Pump cassette Pump roller...
Page 464: ...464 SERVICE 466 SERVICE 3 6 Hach Lange 10 Q1 Q2 QA A A PHOSPHAX sigma 30 48...
Page 465: ...465 11 STATUS 10 C 5 6 6 7 PHOSPHAX...
Page 476: ...476 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Page 477: ...477 Status STATUS SIGMATAX SIGMATAX STATUS SIGMATAX SIGMATAX SIGMATAX STATUS...
Page 479: ...480...