drain hose (LZP 573), which is simply attached from the outside
(
All external hoses should be routed as smooth and bend-free as
possible. The drain hose must be led on the shortest route (maximum
2 m) to a lower-lying and generously dimensioned drain, which even
during the frost season also guarantees an outflow without any backing-
up! A blockage in the drain hose may cause the hoses to spring off
either in the device or outside.
Figure 5 Hose connections
Electrical connections
Current output and limit-value contacts can be assigned by the user
according to requirement. Three PG screw connections (for cable
diameters of 6–9 mm) serve to lead the connection lines on the left
housing side, which are sealed on delivery by a plastic disc.
The device is equipped with a 1.4 m long mains connection cable.
Protection of the device (without refrigerator) occurs through two mains
fuses (T 2.5 A L; 250 V), which are located on the power supply unit
board behind the front panel.
Change of the fuse
Sample feed hose.
Notice! Do not shorten
or lengthen.
Drain hose
D A N G E R
Some circuits in this device are protected by fuses against overvoltages. For
constant protection against risk of fire, replace these fuses only with fuses of the
same type and power rating.
D A N G E R
Defective fuses generally indicate a problem in the device. If fuse errors
repeatedly occur, contact the service department for instructions on the return
procedure for repair purposes. Under no circumstances attempt to repair the
device independently.
N O T I C E
On removal of the power plug (fixed installation of mains supply lead), a suitable
bipolar circuit breaker must be installed!
N O T I C E
Pull out power plug before the opening of the device.
Summary of Contents for PHOSPHAX sigma High Range
Page 5: ...5 II 2 230 V AC 10 50 Hz 145 VA x 540 mm x 600 mm x 400 mm 22 kg...
Page 7: ...7 12 2005 2002 96 Hach Lange Hach Lange...
Page 9: ...9 1 SERVICE 2 3 5 6 3 4 4 3 CEE 10 A STO P 1 2 3 4 5 6 7 5 6...
Page 10: ...10 4 1 2 m SIGMATAX 2 PG LZP 573 5 2 m 5 PG 6 9 mm 1 4 m T 2 5 A L 250 V...
Page 11: ...11 6 SIGMATAX 2 SIGMATAX 2 1 2...
Page 15: ...15 15 PHOSPHAX sigma 1 2 m 5 100 C SERVICE C D 9 48...
Page 16: ...16 3 A C D 6 6 1 2 a SERVICE 18 SERVICE 3 6 Hach Lange 10 30...
Page 17: ...17 Q1 Q2 QA A A PHOSPHAX sigma 11 10 C 5 6 48 6 7 PHOSPHAX...
Page 29: ...29 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Page 32: ...32...
Page 114: ...Deutsch 114...
Page 226: ...Fran ais 226...
Page 282: ...Italiano 282...
Page 370: ...Rom n 370...
Page 452: ...T rk e 452...
Page 456: ...456 2005 8 12 2002 96 EC Hach Lange Hach Lange...
Page 463: ...463 100 C SERVICE C D 9 3 A C D 6 6 1 2 48 Pump cassette Pump roller...
Page 464: ...464 SERVICE 466 SERVICE 3 6 Hach Lange 10 Q1 Q2 QA A A PHOSPHAX sigma 30 48...
Page 465: ...465 11 STATUS 10 C 5 6 6 7 PHOSPHAX...
Page 476: ...476 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Page 477: ...477 Status STATUS SIGMATAX SIGMATAX STATUS SIGMATAX SIGMATAX SIGMATAX STATUS...
Page 479: ...480...