32
Installation
3.7.3 Measurement board
3.8 Measurement alarm relays
Figure 18 Measurement board
Figure 19 Connector J7
Connector J7 (inputs & outputs)
Measurement alarms relays
1.
Common
2.
Output relay 1
3.
Output relay 2
4.
Output relay 3
Analog current (or voltage) outputs
5.
Analog GND
6.
Output 1
7.
Output 2
8.
Output 3
Digital inputs
9.
Hold input. To deactivate the sensor from
a PLC system, connect a dry contact
between J7.9 and J7.12
Note:
It is recommended to use this functionality in
order to extend the sensor lifetime for installations
with a CIP process that can damage the spot.
10.
Not used
11.
Not used
12.
Digital GND
13. to 16.
Not used
Figure 20 Output relays
The three output relays are located on the measurement
board.
They can be individually configured to Normally Open (NO) or
to Normally Closed (NC) by physically moving the jumper on
each relay. On the picture example on the left:
•
Upper relay is set to NC
•
Middle relay is set to NO
•
Lower relay is shown with no jumper
Note:
J14 (upper relay) is relay 1
J18 (middle relay) is relay 2
J19 (lower relay) is relay 3
Summary of Contents for ORBISPHERE K1200
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 39: ...38 Installation...
Page 45: ...44 User Interface...
Page 46: ...45 Section 5 View Menu Figure 34 View menu...
Page 49: ...48 View Menu...
Page 59: ...58 Calibration Menu...
Page 60: ...59 Section 8 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 62: ...61 Inputs Outputs Menu 8 4 Analog outputs Figure 41 Analog outputs menu...
Page 69: ...68 Inputs Outputs Menu...
Page 86: ...85 Section 11 Products Menu Figure 44 Products menu...
Page 89: ...88 Global Configuration Menu...
Page 90: ...89 Section 13 Services Menu Figure 46 Services menu Part 1...
Page 91: ...90 Services Menu Figure 47 Services menu Part 2...