92
Services Menu
13.4 Screen
13.5 Buzzer
13.6 Boards info
13.7 Batteries
13.8 Software download
13.9 End application
Screen
Screen calibration
This screen allows you to adjust the click position corresponding to the
displayed buttons. Use it if ever the sensitive areas are no longer properly
aligned with the buttons on display.
Place the stylus right on the cross when asked and proceed. User will be
asked to press on the screen to accept the new setting. If not, the new
setting is not recorded and no change is made.
Screen brightness
Press the up or down arrow to increase or decrease the screen brightness.
Four levels are pre-set:
Max
,
Comfortable
,
Standard
(default setting) and
Economy
. Press
OK
when finished.
Note:
This can also be called through the brightness icon on the
main display.
Buzzer
Adjust the sounds available on the instrument.
When “screen tap” is enabled, a click sound is heard each time the screen is touched.
The instrument alarm sound can be enabled or disabled to suit the application. The sound type can also
be adjusted. Press the
Test
button to test the adjustment made. Press again to stop.
Boards info
Main board info
For reference, this display gives information on the instrument model,
software version and instrument ID.
Measurement board info
For reference, this display gives information on the sensor measurement
board hardware and software.
Pressing the
Analog outputs corr.
button displays the correction factor that
is applied to the analog outputs.
Pressing the
Amplifiers corr.
button displays the value of the actual
correction factor on the amplifiers.
Sensor parameters
For reference this display gives information on the sensor model and type,
last calibration, settings and behavior.
Batteries
On all instruments this display gives the real time clock battery charge level and voltage.
Note:
A warning message (and icon) is displayed if the battery level becomes too low and needs to be
replaced.
Software download
For Hach technician use only. Used when reloading the software for new versions.
This ends the application. User must stop and restart the instrument to restart the program.
End application
This ends the application. User must stop and restart the instrument to restart the program.
Summary of Contents for ORBISPHERE K1200
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 39: ...38 Installation...
Page 45: ...44 User Interface...
Page 46: ...45 Section 5 View Menu Figure 34 View menu...
Page 49: ...48 View Menu...
Page 59: ...58 Calibration Menu...
Page 60: ...59 Section 8 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 62: ...61 Inputs Outputs Menu 8 4 Analog outputs Figure 41 Analog outputs menu...
Page 69: ...68 Inputs Outputs Menu...
Page 86: ...85 Section 11 Products Menu Figure 44 Products menu...
Page 89: ...88 Global Configuration Menu...
Page 90: ...89 Section 13 Services Menu Figure 46 Services menu Part 1...
Page 91: ...90 Services Menu Figure 47 Services menu Part 2...