![h/p/cosmos Arsalis Gaitway 3D Service Manual Download Page 70](http://html.mh-extra.com/html/h-p-cosmos/arsalis-gaitway-3d/arsalis-gaitway-3d_service-manual_574521070.webp)
gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 66
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
29.
Position 2 is with the gauge axis in between
the rear load cell and the center of the belt.
Position 3 is with the gauge axis close to the
center of the belt.
Position 4 is with the gauge axis in between
the front load cell and the center of the belt.
Position 5 is with the gauge behind the front
load cell.
30.
Ensure that all criterions are passed and
press “Validate”. Otherwise repeat the steps
of a given position that show the largest
errors or repeat all steps from the beginning.
31.
Press “Store & report calibration matrix” to
save it to the FactoryID device flash memory
32.
Attach the calibration report generated to the
calibration summary sheet in the treadmill
logbook.
33.
Perform on-site checkout procedure.
END OF PROCEDURE