gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 37
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
46.
Measure impedance between the backshell
of the power connector on the force amplifier
and the treadmill PE.
Ensure that no cable is connected to the
force amplifier.
R < 1ohm
47.
Measure impedance of the power cable
shielding (on the connector backshell) to the
treadmill PE.
R < 1ohm
48.
Measure the impedance of the power cable
ground (pin3) and 12V (pin1) to the treadmill
PE.
R
gnd
< 1ohm
R
12V
>100ohm
49.
Ensure that test area is clear from tools and
that the treadmill can be powered safely.
Then, connect the treadmill mains plug.
50.
Place an emergency stop plug on the safety
arch plug if necessary.
Bridge plug XLR, cos11685