![background image](http://html.mh-extra.com/html/h-p-cosmos/arsalis-gaitway-3d/arsalis-gaitway-3d_service-manual_574521039.webp)
gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 35
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
43.
Tie the power cable to the cables inside the treadmill such as
on the picture.
3x Tie-wrap
44.
Insert the terminal ring of the power cable on
the Earth terminal post of the treadmill and
re-screw all terminal rings using the washers
and nut supplied with the treadmill.
Torque =
6Nm