gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 52
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
70.
Connect the 4 load cell cables to the
amplifier.
71.
Adjust the length of the four load cell cables.
Roll the cable excess into the cable trays on
the foot plates. Close the cable trays.
Tighten the tie wraps on the load cell cables.
72.
Inspect that the load cell cables cannot
touch any sharp part.