gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 59
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
7.
Turn on the power then wait for 1hour for
temperature stabilization before starting the
calibration measurements.
8.
Start Gaitway-3D[Service] software from the
USB stick provided with the calibration
equipment and report software version.
Version:
Click on logos to get the software version.
9.
In gaitway-3D[Service] software,
select:
“Signals noise check” and verify system is
operational including the calibration gauge.
For router and PC Ethernet configuration
10.
In the gaitway-3D[Service] software, select
“Speed sensor calibration”.