![h/p/cosmos Arsalis Gaitway 3D Service Manual Download Page 36](http://html.mh-extra.com/html/h-p-cosmos/arsalis-gaitway-3d/arsalis-gaitway-3d_service-manual_574521036.webp)
gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 32
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
39.
Instrumentation power cable installation
Pass the power cable through the RIGHT
grommet of the treadmill front. Then pass
the cable end through the hole of the main
rib of the motor compartment with the other
cables of the treadmill.
Leave 10 ± 1 cm of cable from the grommet
to end of molded backshell of M12 connector
for bend relief.