gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 100
17. gaitway-3D amplifier fuse replacement checklist
Date:
Name:
Time:
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
1.
Test Readiness Review:
Check
implementation
of
safety
precautions
Identify item name, identification number
and serial number.
Device name, type:
Device s/n:
Instrumentation s/n:
2.
Disconnect main plug.
3.
Perform a visual inspection of the treadmill,
treadmill amplifier and all visible cables for
visible damage.
No visible
damage ?
4.
Disconnect all accessories connected on the
top panel of the amplifier.
5.
Disconnect the power cable.