SECTION 10 PERFORMANCE SPECIFICATIONS
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 10-5
In addition to the primary outputs above other waveforms are available as described below:
Blended Waveforms, T1 & T2:
T1
–
A blended output with 50% at PK2, 50% at DES, oscillates at 30 Hz or 500 Hz
depending on instrument type.
T2
-
A blended output with 50% at PK3, 50% at DES, oscillates at 30 Hz or 500 Hz
depending on instrument type.
Vapor Pulse Waveforms, VP3, VP2 & VP1:
Pulsed coagulation outputs with a limited maximum output voltages. During the off periods, no RF
output occurs. Repetition rates for each VP waveform will be constant but may vary with different
instrument types.
Voltage limits:
VP3 -
120Vrms
VP2 -
100Vrms
VP1 -
73Vrms
The output power is defined as follows:
CycleTime
OnTime
PeakPower
er
AveragePow
Where the Peak power is constant, the cycle time (the time for a complete on/off period) is preset
and the On/Off ratio is varied dependent on set power.
SECTION 10
PERFORMANCE SPECIFICATIONS