SECTION 4 GENERAL INFORMATION
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 4-4
When the desired mode is displayed, a further press on the Mode / Menu button will
complete the selection
Fig 4.4 Screen for coag waveform selection
Socket Selection
The generator permits two instruments to be fitted simultaneously, via the two 5-way
connectors.
Only one socket can deliver RF at any one time. When the generator is switched on from
the standby state, it initially operates an automatic socket selection mechanism, and assists
primary connection by activating whichever socket first has an instrument attached. The
instrument connected to the active socket is indicated on the display. Thereafter socket
selection can only be altered manually, by pressing the black mode footswitch, ensuring
that the surgeon always has control over which instrument is activated.
Fig 4.5 Footswitch
Footswitch - Blue Pedal
The blue footswitch pedal is used to administer desiccate and VP output waveforms. The
output will be present while the footswitch is held down.
Footswitch – Yellow Pedal
The yellow footswitch pedal is used to administer PK (cut) and ThermoKinetic waveforms.
The output will be present while the footswitch is held down.
Footswitch – Black Mode Footswitch
The black mode footswitch enables a rapid change between PK instruments.
Select Mode
DES
Black Mode
Footswitch
SECTION 4
GENERAL INFORMATION