SECTION 14 EMC TABLES
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The SuperPulse Generator is intended for use in the electromagnetic environment specified below. The
customer or the user of the SuperPulse Generator should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The SuperPulse Generator uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies
The SuperPulse Generator is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Page 14-1
SECTION 14
EMC TABLES
Summary of Contents for PlasmaKinetic SuperPulse
Page 2: ......