SECTION11 ERROR AND FAULT CODES
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 11-2
200
10
INTERNAL FAILURE
[F]
PWM signal: shorted high
200
11
INTERNAL FAILURE
[F]
PWM signal: shorted low
200
12
INTERNAL FAILURE
[F]
SYNC signal: shorted high
200
13
INTERNAL FAILURE
[F]
SYNC signal: shorted low
200
14
INTERNAL FAILURE
[F]
ENERGY signal: stuck high
200
15
INTERNAL FAILURE
[F]
IOUT signal: stuck high (comparator in)
200
16-17
Not used
---
----
200
18
INTERNAL FAILURE
[F]
CUT/COAG signal: stuck high
200
19
INTERNAL FAILURE
[F]
CUT/COAG signal: stuck low
200
20
INTERNAL FAILURE
[F]
CLAMP signal error (DAC output)
200
21
INTERNAL FAILURE
[F]
BOOST signal error (DAC output)
200
22
INTERNAL FAILURE
[F]
PEAKSET signal error (DAC output)
200
23
INTERNAL FAILURE
[F]
PEAK signal error: stuck high (comparator in)
200
24
INTERNAL FAILURE
[F]
RF_DET signal error: stuck low
200
25
INTERNAL FAILURE
[F]
OVERDOSE signal error: permanently ON
200
26
INTERNAL FAILURE
[F]
ENERGY signal error: stuck low
200
27
Not used
---
----
200
28
INTERNAL FAILURE
[S]
Temperature monitor inoperative
200
29
------
----
Reserved for audio fault detection
200
30
Not used
----
-----
200
31
INTERNAL FAILURE
[F]
CURRLIM signal error (DAC output)
200
32
INTERNAL FAILURE
[F]
VOLTLIM signal error (comparator input)
200
33
INTERNAL FAILURE
[F]
BUSVOLTS signal error (analogue input)
200
34
INTERNAL FAILURE
[F]
Incorrect PK RF board installed
200
35
INTERNAL FAILURE
[F]
RF relay 1 (socket) not operating
200
36
INTERNAL FAILURE
[F]
RF relay 2 (poles) not operating
200
37
INTERNAL FAILURE
[F]
SP board RF_ACTIVE stuck on
200
38
INTERNAL FAILURE
[F]
SP board Output relay non-functional
200
39
INTERNAL FAILURE
[F]
SP board BUS relay non-functional
200 40-2
Not
used
---
-----
200
43
INTERNAL FAILURE
[F]
CPU POST failure
200
44
RAM
[R]
RAM test failure [POST check]
200
45
INTERNAL FAILURE
[F]
Crystal failure [POST check]
200
47
INTERNAL FAILURE
[F]
Analogue reference failure [POST check]
200
48
INTERNAL FAILURE
[F]
CLAMP_SET failure [POST check]
200 49
INTERNAL
FAILURE [F]
BOOST_SET failure [POST check]
200
50
INTERNAL FAILURE
[F]
CURRLIM_SET failure [POST check]
200
51
INTERNAL FAILURE
[F]
PEAK_SET failure [POST check]
200
52
INTERNAL FAILURE
[F]
COAG_CUT failure [POST check]
200
53
INTERNAL FAILURE
[F]
PK_CUT_COAG failure [POST check]
200
54
INTERNAL FAILURE
[F]
PK_SKT_SET failure [POST check]
200
55
INTERNAL FAILURE
[F]
SUPERPULSE_RELAY failure [POST check]
200
56
INTERNAL FAILURE
[F]
NO_RF failure [POST check]
SECTION 11
ERROR AND FAULT CODES
Summary of Contents for PlasmaKinetic SuperPulse
Page 2: ......