SECTION 2 PATIENT AND OPERATING ROOM SAFETY
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
• Keep the accessory cables away from cables from other electrical equipment. Electrical
currents may be induced in the other equipment causing unintended effects.
• Do not use a monopolar generator/accessories simultaneously with the SuperPulse
generator. Activation of a monopolar generator/accessories may cause interference with
the SuperPulse generator resulting in user message changes on the display. Before
proceeding with surgery, confirm proper power settings are displayed on the generator.
Ensure the appropriate output setting is enabled for the desired surgical outcome.
• Provide as much separation as possible between the generator and other electronic
equipment (such as monitors). When activating the generator, unintended electromagnetic
coupling may cause interference with the other equipment.
• Should any unintentional effects appear upon other equipment when using the generator,
repositioning the generator, the connecting leads or other equipment may alleviate the
problem. It may also help to use different mains supply sockets for any affected
equipment
Page 2-6
SECTION 2
PATIENT AND OPERATING ROOM SAFETY
Summary of Contents for PlasmaKinetic SuperPulse
Page 2: ......