SECTION 6 DURING SURGERY
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 6-3
6F. Changing Accessories Between Procedures
Section 8 describes the disconnection of the connector cable and instrument.
Once the connector cable is disconnected, the generator will revert to the display showing
“Connect PK Cable”
.
The generator can be left in this Idle state between procedures.
CAUTION
If the generator is maintained in this Idle state between procedures and
the same PK instrument type is employed for the next procedure, then
any adjustments made to the output settings during the previous
procedure will be remembered and will supersede the default settings.
Inserting a new PK instrument type will reset the generator to the default settings for that
particular type. Any adjustments to the PK Generator settings made during the previous
procedure will be overridden.
Switching the power off will clear all prior PK output adjustments.
SECTION 6
DURING SURGERY