PCS-1000 User Manual
90
AC Characteristics
True RMS AC
Voltage
Range
Frequency
1 Year
23˚C ± 5˚C
Temperature
Coefficient/˚C
200.0000 mV
45 Hz - 2 kHz
2 kHz - 10 kHz
10 kHz - 20 kHz
0.5 + 0.05
1.0 + 0.05
2.0 + 0.10
0.005 + 0.005
2.000000 V
0.005 + 0.005
20.00000 V
0.005 + 0.005
200.0000 V
0.005 + 0.005
600.000 V
0.005 + 0.005
Accuracy specification : ± (% of r % of range)
True RMS AC
Current
Range
Frequency
1 Year
23˚C ± 5˚C
Temperature
Coefficient/˚C
(AC+DC)
30.00000 mA
45 Hz - 2 kHz
2 kHz - 10 kHz
0.5 + 0.05
1.0 + 0.05
0.03 + 0.006
300.0000 mA
0.03 + 0.006
3.000000 A
0.03 + 0.006
30.00000 A
45 Hz - 400 Hz 0.5 + 0.05
0.03 + 0.006
300.0000 A
0.03 + 0.006
Accuracy specification : ± (% of r % of range)
Current Monitor Accuracy
Range
Shunt Value
DC Accuracy
AC Accuracy
≦
400 Hz
Max Input
DC/AC rms
30.00000 mA 10
Ω
0.01%
0.1%
30 mA
300.0000 mA 1
Ω
0.01%
0.1%
300 mA
3.000000 A
0.1
Ω
0.01%
0.1%
3 A
30.00000 A
0.01
Ω
0.01%
0.1%
30 A
300.0000 A
0.001
Ω
0.02%
0.1%
300 A
Summary of Contents for PCS-1000
Page 52: ...PCS 1000 User Manual 52...