COMMUNICATION INTERFACE
79
Example
SYST:OPER?
>256
Indicates that bit 8 has been latched.
STATus:PRESet
Set
Description
Resets the Standard Event Enable Register, the
Questionable Data Enable Register and the
Standard Operation Enable Register to their
default state.
Syntax
STATus:PRESet
STATus:QUEStionable:CONDition
Query
Description
Returns the contents of the Questionable Data
Condition Register.
Bit
Bit weight Description
0
1
Volt Overload
1
2
Current Overload
2~15
~
Not used
Query Syntax
STATus:QUEStionable:CONDition?
Return Parameter <NR1>
0~65535: Returns the bit weight of the
Questionable Data Condition Register.
Query Example
STAT:QUES:COND?
> 1
Indicates there was a voltage overload.
Summary of Contents for PCS-1000