PCS-1000 User Manual
74
SYSTem:ERRor
Query
Description
Queries the error queue. Error messages are stored
in FIFO order. Up to 20 error messages are stored
in the error queue. The first error message that is
stored is the first message that is returned. Each
time a message is returned it is also cleared from
the queue. When the error queue is queried and
there are no error messages, 0, "No error" will be
returned. If the error queue is full (20 messages)
and an error occurs, the last-stored error message
will be overwritten with the
-350,"Error queue overflow" message. This
message will remain, and no additional messages
will be stored until it is cleared.
See page 87 for a list of the error messages.
Query Syntax
SYSTem:ERRor?
Return Parameter <string>
Returns the next error message in the
error queue.
Query Example
SYST:ERR?
> 0, "No error."
Returns no error in the error queue.
SYSTem:LOCal
Set
Description
Returns the unit back to local mode. This
command will enable all panel keys that may have
been locked.
Syntax
SYSTem:LOCal