COMMUNICATION INTERFACE
69
Sense Commands
[SENSe:]CURRent:RANGe .......................................... 69
[SENSe:]CURRent:DC:AVERage:COUNt .................... 70
[SENSe:]CURRent:AC:AVERage:COUNt .................... 70
[SENSe:]VOLTage:RANGe .......................................... 70
[SENSe:]VOLTage:DC:AVERage:COUNt .................... 71
[SENSe:]VOLTage:AC:AVERage:COUNt .................... 72
[SENSe:]CURRent:RANGe
Set
Query
Description
Sets or queries the current range.
Syntax
Query Syntax
[SENSe:]CURRent:RANGe {<Range>|AUTO}
[SENSe:]CURRent:RANGe?
Parameter /
Return Parameter
<Range>
Current range <NRf>: 0.00000001
~
305
Sets the current range in amps. The unit
will automatically choose the closest
range that is programmed.
AUTO
Sets the range to AUTO; Only applicable
for the ≤3A ranges.
Autorange is not supported for the 30A
and 300A ranges.
Example
CURR:RANG AUTO
Sets the current range to AUTO.
Note
The range that is returned is the base unit. See the
table below:
Unit
Current Range
100
300A
10
30A
1
3A
.1
300mA
.01
30mA