COMMUNICATION INTERFACE
65
CONFigure:VOLTage:AC
Set
Description
This command will set the voltage mode to AC
and set the ACV range. If the range is not specified
then it will not be changed.
Syntax
CONFigure:VOLTage:AC [<Range> | AUTO]
Parameter
<Range>
Voltage range <NRf>: 0.0000001~630
The unit will automatically be set to the
closest range.
AUTO
Autoset
Example
CONF:VOLT:AC 20
Sets the voltage mode to AC and the ACV range to
20V.
Example
CONF:VOLT:AC
Sets the voltage mode to AC. The range stays the
same.
CONFigure:AVERage:MODE
Set
Query
Description
This command will set or query the average mode.
Syntax
Query Syntax
CONFigure:AVERage:MODE {0|1,TOTAL|SHIFT}
CONFigure:AVERage:MODE?
Parameter
0, TOTAL
Total mode
1, SHIFT
Shift mode
Return Parameter Total
Total mode
Shift
Shift mode
Example
CONF:AVER:MODE 0
Sets the average mode to Total mode.