GDM-8352 User Manual
144
DC Voltage
Range
[1]
Resolution
Full Scale
Accuracy
200.000mV
0.001mV
239.999
0.012% + 8
2.00000 V
0.00001V
2.39999
0.012% + 5
20.0000 V
0.0001V
23.9999
0.012% + 5
200.000 V
0.001V
239.999
0.012% + 5
1000.00 V
[2]
0.01 V
1020.00
0.012% + 5
[1] When the input value exceeds the full scale of the selected range, the display
will show -OL- (over load) on the display.
[2] The specifications are guaranteed to an input voltage of 1000V. A beeping
alarm will go off when the input voltage is higher than 1000V.
DC Current
Range
[1]
Resolution
full scale
Accuracy
20.0000mA
0.0001mA
23.9999
0.05% + 15
200.000mA
0.001mA
239.999
0.05% + 5
1.00000A
0.00001 A
1.19999
0.2% + 5
10.0000A
[2]
0.0001 A
11.9999
0.2% + 5
[1] When the input value exceeds the full scale of the selected range, the display
will show -OL- (over load) on the display.
[2]The specifications are guaranteed to an input of 10A. A beeping alarm will go
off when the input value is higher than 10A.
Diode
Range
Resolution
Maximum reading
Accuracy
6V
0.0001 V
5.9999V
0.05% + 15
*The diode test Current is 1mA.
*Accuracy specifications are for the voltage measured at the input terminals
only.
Continuity
Range
Resolution
Maximum reading
Accuracy
2000.00
Ω
0.01
Ω
2399.99
0.05% + 5
* Without REL function, add 0.2
Ω
additional error.