GDM-8352 User Manual
150
AC Current (AC Coupling Mode/AC + DC Coupling Mode)
Measurement method: Current to the fuse and current shunt, AC
coupled true RMS measurement (measures the AC component
only).
Crest factor: Maximum of 3 at full scale.
Rate
Range
Accuracy
20 Hz to 45 Hz 45 Hz to 2 kHz 2 kHz to 10kHz
Med
20.0000mA
1.5% + 400
0.5% + 400
2% + 800
200.000mA
1.5% + 120
0.5% + 120
2% + 300
1.00000A
1.5% + 120
0.5% + 120
2% + 300
10.0000A
2% + 120
1% + 120
-
Fast
20.0000mA
-
0.5% + 500
2% + 1000
200.000mA
-
0.5% + 200
2% + 500
1.00000A
-
0.5% + 200
2% + 500
10.0000A
-
1% + 200
-
Rate
Additional Rate Error Count
Med
50
Fast
500
Shunt resistance
Range
SHUNT
Burden voltage
20mA
1.1
Ω
<
0.08V
200mA
1.1
Ω
<
0.8V
1A
0.1
Ω
<
0.8V
10A
0.01
Ω
<
0.6V