COMMAND OVERVIEW
119
[SENSe:]AVERage:COUNt
Sets the digital filter average count.
Parameter: <NR1> (2~320)|MIN|MAX
Example: SENS:AVER:COUN 100
Sets the digital filter average count to 100.
[SENSe:]AVERage:COUNt?
Returns the current digital filter average count.
Return parameter: <NR1>(+002 ~ +320)
[SENSe:]AVERage:WINDow
Sets the digital filter window.
Parameter: 0.01|0.1|1|10|0 (none)
Example: SENS:AVER:WIND 0.1
Sets the digital filter window to 0.1%.
[SENSe:]AVERage:WINDow?
Returns the current digital filter window value.
Return parameter: 0.01, 0.1, 1, 10, NONE
[SENSe:]AVERage:STATe
Turns the digital filter on or off.
Parameter: ON|OFF
Example: SENS:AVER:STAT ON
Turns the digital filter on.
[SENSe:]AVERage:STATe?
Returns the state of the digital filter (on or off).
Return parameter: 0|1, 0=OFF, 1=ON