COMMAND OVERVIEW
113
Measure Commands
MEASure:VOLTage:DC?
Returns the DC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:DC?
> +0.10348E-01
Returns the DC voltage measurement as 0.010348 V.
MEASure:VOLTage:AC?
Returns the AC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:AC?
> +0.09020E-01
Returns the AC voltage measurement as 0.009020V.
MEASure:VOLTage:DCAC?
Returns the DC+AC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:DCAC?
> +0.10123E-01
Returns the DC+AC voltage measurement as 0.010123V.
MEASure:CURRent:DC?
Returns the DC current measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:CURR:DC?
> +0.00703E-02
Returns the DC current measurement as 0.0703 mA.