COMMAND OVERVIEW
133
Interface Commands
SYSTem:LOCal
Enables local control (front panel control) and disables remote
control.
SYSTem:REMote
Enables remote control and disables local control (front panel
control). Local control can be recalled by pressing the 2ND or
local button.
SYSTem:RWLock
Enables remote control and disables local control (front panel
control). Once this command has been issued, pressing the 2ND
or local buttons will not return the user to local control. The only
way to return to local mode is to issue the SYSTem:LOCal
command.
IEEE 488.2 Common Commands
*CLS
Clears the Event Status register (Output Queue, Operation Event
Status, Questionable Event Status, Standard Event Status)
*ESE?
Returns the ESER (Event Status Enable Register) contents.
Example: *ESE?
>130
Returns 130. ESER=10000010