GDM-8352 User Manual
56
Using Dual Measurement Mode
Procedure
1.
Choose one of the basic measurement functions
from the table above to set the measurement
mode for the primary display.
For example, press DCV to set the first display
to DCV measurement.
2.
To set a measurement mode for the second
display, press the 2ND key and then select the
second measurement mode.
For example, press 2ND, SHIFT, ACV(ACI) to
select ACI measurement for the second display.
Display
Indicators for 1
st
measurement
2
nd
measurement
and unit
Indicators for 2
nd
measurement
1
st
measurement
and unit
REAR
GPIB
CV
CC
SWEEP
FILT
RCL
STO
COMP
2ND
dBm
EXT
SHIFT
ERR
LIMIT
AUTO
MATH
4W
2W
DC
AC
A
V
m
m
Hz
W
k
M
W
S
F
C
m
dB
%
A
V
m
m
Hz
W
k
M
W
S
F
C
RMT
MAX
MIN
REL
HOLD
S
M
F
AUTO
4W
2W
DC
AC
n
μF
Note
When in the dual measurement mode, the
reading on the secondary display will be
reduced to 4½ digits by dropping the last digit;
the full 5½ digit measurement reading is still
retained, just not displayed. The full 5½ digit
reading can still be read remotely.