GDM-8352 User Manual
108
CONFigure:FREQuency
Sets measurement to Frequency on the first display and specifies
the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:FREQ MAX
Sets the frequency measurement range to max.
CONFigure:PERiod
Sets measurement to Period on the first display and specifies the
range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:PER
Sets the DMM to period measurement using the autorange.
CONFigure:CONTinuity
Sets measurement to Continuity on the first display.
Parameter: None
CONFigure:DIODe
Sets measurement to Diode on the first display.
Parameter: None
CONFigure:TEMPerature:TCOuple
Sets measurement to Temperature thermocouple on the first
display.
Parameter: [None] | [Type(J | K | T)]
Example: CONF:TEMP:TCO J
Sets the measurement mode to TCO with a type J sensor.