OPERATION
33
Select the Voltage Range
The voltage range can be set automatically or manually.
Auto Range
To turn the automatic range selection On/Off,
press the AUTO key.
Manual Range
Press the Up or the Down key to select the
range. The AUTO indicator turns Off
automatically. If the appropriate range is
unknown, select the highest range.
Selectable Voltage
Ranges
Range
Resolution
Full scale
200mV
1
μ
V
239.999mV
2V
10
μ
V
2.39999V
20V
0.1mV
23.9999V
200V
1mV
239.999V
750V (AC)
10mV
765.00V
1000V (DC)
10mV
1020.00V
Note
For further details, please see the specifications on
page 143.