51
Conducted RF
lEC 61000-4-6
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
6 Vrms
ISM
frequency
band
3 V/m
80 MHz to 2.7
GHz
3V
6V
3V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of
the models Ai-Motor
、
MotoPex,
including cables, then the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d=1.2×P1/2
d=2×P1/2
d=1.2×P1/2 80 MHz to 800 MHz
d=2.3×P1/2 800 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter
In watts (W) according to the
transmitter manufacturer and d
Is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
a
should be less than the
compliance level in each
frequency range
.b Interference may occur In the
vicinity of equipment marked
with
t
he following symbol:
NOTE I At 80 MHz end 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.