28
Motor combined canal
measurement function mode
When using the motor
combined with the integrated
apex locator function, connect
the lip hook to the socket
(white) on the measurement
cable to monitor working length
during the root canal
preparation, keep the black
socket idle.
The canal length indicator bar
will be displayed on the screen
(more information about canal
length indicator bar, please see
chapter 3. 2 Screen display)
Setting parameters of automatic
functions as needed, such as
Apical Action, Auto Start, etc
(more information about
automatic functions, please see
chapter 4.3
Parameter setting)
Connection testing
It is strongly recommended to
check the connection testing every
time before use.
Touch the lip hook with the
file inserted to the contra-angle
and check that all the bars on the
display on the screen light up,
and the rotation should be
reversed. The measurement cable
or contra-angle should be
replaced if these functions are
not noticed.