42
t
o the rinsing connection of the washer-disinfector.
3.
Start the program.
4.
After the program is finished, remove the product from the washer-
disinfector, inspect (refer to the section "Inspection and Maintenance") and
the packaging (refer to chapter "Packaging"). Repeat the process if necessary
(refer to section "Drying").
Notes:
a)
Before use, you must carefully read the operating instructions provided
by the equipment manufacturer to familiarize yourself with the disinfection
process and precautions.
b)
With this equipment, cleaning, disinfection and drying can be carried
out together.
c)
Cleaning: (c1) The cleaning procedure should be suitable for the product
to be treated. The flushing period should be sufficient (5-10 minutes). Pre-
wash for 3 minutes, wash for another 5 minutes and rinse it twice with each
rinse lasting for 1 minute. (c2) when washing, the water temperature should
not exceed 45 °C, to prevent solidification of the protein which is difficult to
remove. (c3) freshly prepared solutions such as pure water, distilled water,
deionized water or multi-enzyme solution, etc. can be used. (c4) During the
use of cleaner, the concentration and time provided by the manufacturer must
be followed. The used cleaner is neodisher MediZym (Dr. Weigert).
d)
Disinfection: (d1) Direct use after disinfection: temperature ≥ 90 ° C,
time ≥ 5 min or A0 ≥ 3000;
Sterilize it after disinfection and use: temperature ≥ 90 ° C, time ≥1 min or A0
≥ 600
(d2) For the disinfection here, the temperature is 93 ° C, the time is 2.5 min,
and A0>3000
e)
Only distilled or deionized water with a small number of
microorganisms (<10 cfu/ml) can be used for all rinsing steps. (For example,
pure water that is in accordance with the European Pharmacopoeia or the
United States Pharmacopoeia).
f)
After cleaning, the chemical residue should be less than 10mg / L.
g)
The air used for drying must be filtered by HEPA.
h)
Regular maintenance and inspection of the disinfector should be
performed.